Comparison of Phase Angle Using Bioelectrical Impedance Analysis According to Device Type and Examination Posture
Yang, J.; Kim, J.; Chun, B. C.; Lee, J.-m.
Show abstract
BackgroundBioelectrical impedance analysis (BIA) is gaining popularity as a body composition assessment tool for patients. Although it has been studied and validated in different populations, age groups, and clinical settings, including critically ill patients, there are concerns about BIA reproducibility and reliability in different device types and postures. This study aimed to evaluate the reliability of BIA according to different devices, postures, and lead types. MethodsCross-sectional observational data of 74 healthy volunteers (32 women and 42 men) were collected. We used two types of devices, three types of postures (standing, sitting, and lying), and two lead types (clamp lead and adhesive lead) to measure the whole-body phase angle (phA) at a single frequency of 50 kHz. The measurements were validated using the intraclass correlation coefficient (ICC) and Bland-Altman plot analysis. ResultsAll phA measurements recorded with two types of devices, three different postures, and two types of leads were equivalent to each other (mean ICC = 0.9932, 95% confidence interval (CI) 0.9905-0.0053). The average mean difference in phA was 0.31 (95% CI 0.16- 0.46). The largest phA value was measured using BWA with an adhesive-type lead in the lying position. There were no differences between standing and sitting positions. ConclusionThis is the first study to show the consistency and reliability of BIA in measuring phA using different devices, lead types, and postures. This could provide the confidence that BIA can be used in various clinical settings.
Matching journals
The top 2 journals account for 50% of the predicted probability mass.
Similar papers in this journal
- Estimation equation of limb lean soft tissue mass in Asian athletes using bioelectrical impedance analysis 95%
- Tongue thickness measured by ultrasonography is associated with tongue pressure in the Japanese elderly 93%
- Comparison of raw accelerometry data from ActiGraph, Apple Watch, Garmin, and Fitbit using a mechanical shaker table 93%
Similar papers in this journal
- Protocol for a prospective cohort study on the feasibility of application of nutritional ultrasound in the diagnosis and follow-up of patients with nutritional risk at hospital discharge: Study on body composition and function (DRECO) 94%
- Wearable Sensing in Eating Episode Monitoring: An Updated Systematic Review Protocol 91%
- Cohort profile: The Westlake BioBank for Chinese (WBBC) pilot cohort: a prospective study for the late adolescence 91%
Similar papers in this journal
- A sweat lactate sensor for detecting anaerobic threshold in patients with heart failure: a prospective clinical trial (LacS-001) 93%
- Ambulatory physiological measures obtained under naturalistic urban mobility conditions have acceptable reliability 93%
- Effects of Electrocardiograms QRS Detection Algorithms in Heart Rate Variability Metrics 92%
Similar papers in this journal
- Amplitude-integrated EEG amplitudes differ between wakefulness and sleep states in children 89%
- Noninvasive optical monitoring of cerebral hemodynamics in a preclinical model of neonatal intraventricular hemorrhage 88%
- A method for measuring mitochondrial DNA copy number in pediatric populations 88%
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.