Evaluation of the Analytical Performance of Hemoglobin Analysis on the Point-of-Care Testing Device HemoCue Hb 801
Olesen, S. V.; Nielsen, L. K.; Gredal, L.
Show abstract
IntroductionEvaluating the accuracy and imprecision of Point-of-Care Testing (POCT) devices is an important prerequisite for ensuring proper diagnosis. The objective of this study was to assess the reliability of POCT hemoglobin (Hb), conducting a method comparison study on HemoCue Hb 801 and Sysmex XN 9000 as reference. MethodsThirty EDTA whole blood samples previously tested for Hb on Sysmex XN 9000 were also analysed on HemoCue Hb 801 within the measurement range 6.4-21.9 g/dL. Data were assessed by linear regression and correlation with a 95% confidence interval (CI) for the slope and intercept, a paired Students t-test, bias calculations and a Bland-Altman plot. Twelve independent measurements on the same sample material determined the imprecision of HemoCue Hb 801. ResultsWe found a strong correlation coefficient (r2) of 0.992 between Hb measured on HemoCue Hb 801 and Sysmex XN 9000 with a slope of 0.984 (CI: 0.956 to 1.021) and an intercept of 0.181 (CI: -0.283 to 0.544). The paired Students t-test indicated that the mean Hb differences between the methods were not statistically significant (p=0.84). Bias (-0.01 g/dL) did not deviate from the allowed maximum absolute bias ({+/-}0.35 g/dL). The values were within the 95% limits of agreement ({+/-}1.40 g/dL) for the differences. The coefficient of variation was 1.0%. ConclusionThere was a strong correlation and good agreement between Hb measured on HemoCue Hb 801 and Sysmex XN 9000, as well as an acceptable bias and imprecision, supporting a good accuracy.
Matching journals
The top 2 journals account for 50% of the predicted probability mass.
Similar papers in this journal
- Comparison of haemoglobin concentration measurements using HemoCue-301 and Sysmex XN-Series 1500: a survey among anaemic Gambian infants aged 6-12 months 95%
- Optimization Of Dried Blood Spot Assays For Hepatitis B Virus Surface Antibody Quantification 95%
- EDTA tubes are suitable for insulin and C-peptide measurement in resource-limited settings and can be stored at room temperature for up to 24 hours 94%
Similar papers in this journal
- A simple, cost-effective and extraction-free molecular diagnostic test for sickle cell disease in noninvasive buccal swab specimen for a limited-resource setting 94%
- Reducing the Cost of Rapid Antigen Tests through Swab Pooling and Extraction in a Device 91%
- Analytical Validation of MyProstateScore 2.0 91%
Similar papers in this journal
Similar papers in this journal
- Prediction of plasma volume and total hemoglobin mass with machine learning 94%
- Erythropoietin production?by the kidney and the liver in response to severe hypoxia evaluated by Western blotting with deglycosylation 91%
- Void spot assay visualization optimization for use of Void Whizzard in rats (Rattus norvegicus) 88%
Similar papers in this journal
- Multispectral Imaging for MicroChip Electrophoresis Enables Point-of-Care Newborn Hemoglobin Variant Screening 96%
- High-sensitive detection and quantitation of thyroid-stimulating hormone (TSH) from capillary/fingerstick and venepuncture whole-blood using fluorescence-based rapid lateral flow Immunoassay (LFIA) 92%
- Retrospective and prospective studies evaluating the performance of the SARS-Cov-2 “AQ+ COVID-19 Ag Rapid Test” from InTec on symptomatic and non-symptomatic patients 91%
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.