Intraluminal pressure in central venous hemodialysis catheters during power injection of contrast media
White, N. A.; van der Molen, A. J.; Limpens, R. W.; Maas, J.; van der Bogt, K. E.; Horeman, T.; Rotmans, J. I.
Show abstract
BackgroundCentral venous catheters (CVCs) provide direct access to the central circulatory system, commonly used in hemodialysis and intensive care units for drug administration. Although uncertified for the procedure, CVCs are sometimes used for power injection of contrast medium (CM) during CT scans to avoid peripheral intravenous catheter placement. Previous studies suggest this practice is safe, but incidents have been reported. This study aims to measure intraluminal pressure during CM injection through CVCs and assess its impact on the luminal surface to guide responsible clinical use. MethodsStrain gauges were applied to the exterior walls of four samples from three different types of unused CVCs. These gauges measured material deformation due to intraluminal pressure during CM injections at rates of 4.5 mL/s and 8 mL/s, each performed five times. Strain data were calibrated against known pressures in a static system. The CVCs were then either pressurized until bursting or subjected to microscopic analysis of their luminal surfaces. ResultsIntraluminal pressures measured (97-545 kPa or 14-79 PSI) were below the burst pressure (779-1248 Kpa or 113-181 PSI) in all instances. Strain regression analysis shows a statistically significant (p<0.05) trend over 10 injections in almost all CVCs tested, indicating material fatigue. Surface microscopy revealed surface micro-cracks from repeated injections, suggesting material damage. ConclusionsThe intraluminal pressures from power injections of CM are sufficiently low to prevent CVC bursting. While incidental use for CM injection appears safe, repeated use may cause material damage
Matching journals
The top 5 journals account for 50% of the predicted probability mass.
Similar papers in this journal
- Reliability and Performance of the IRRAflow® System for Intracranial Lavage and Evacuation of Hematomas - A Technical Note 97%
- A coupled atrioventricular-aortic setup for in-vitro hemodynamic study of the systemic circulation: Design, Fabrication, and Physiological relevancy 93%
- Variability of Tissue Mechanical Response in Sus Domesticus Porcine Models from in vivo to ex vivo Conditions 92%
Similar papers in this journal
- Multimodal Thrombectomy Device for Treatment of Acute Deep Venous Thrombosis 94%
- A Niclosamide-releasing hot-melt extruded catheter prevents Staphylococcus aureus experimental biomaterial-associated infection 94%
- Hemodynamics with Mechanical Circulatory Support Devices Using a Cardiogenic Shock Model 93%
Similar papers in this journal
- Switching the left and the right hearts: A novel bi-ventricle mechanical support strategy with spared native single-ventricle 93%
- Effect of Sinotubular Junction Size on TAVR Leaflet Thrombosis: A Fluid-structure Interaction Analysis 92%
- Urological benchtop and in silico models validated by human penile tissue inflation tests 92%
Similar papers in this journal
- Intracranial pressure elevation alters CSF clearance pathways 90%
- Human intracranial pulsatility during the cardiac cycle: a computational modelling framework 88%
- Modeling CSF circulation and the glymphatic system during infusion using subject specific intracranial pressures and brain geometries 87%
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.