A Pilot Study of Train-of-Four and Post-Tetanic Count Monitoring with the TetraGraph Electromyograph Compared to the TwitchView Monitor Electromyograph
Bussey, L.; Jelacic, S.; Togashi, K.; Bowdle, A.
Show abstract
BackgroundQuantitative twitch monitoring is recommended for avoiding residual neuromuscular blockade. Electromyograph twitch monitors are a form of quantitative twitch monitoring. The TwitchView electromyograph has been previously validated against "gold standard" mechanomyography, and may serve as a comparator for other monitors. We have previously shown that the GE electromyograph monitor overcounted twitches, frequently misinterpreting noise as a twitch. This is a pilot study to evaluate the performance of the TetraGraph electromyograph in comparison to the TwitchView electromyograph. MethodsTwitchView and TetraGraph electrodes were applied to opposite arms of patients prior to induction of anesthesia. Post-tetanic count, train-of-four count and train-of-four ratio were then measured approximately every 5 minutes during an unrestricted general anesthetic. Measurements were not made for 10 minutes following neuromuscular blocking drugs or reversal agents. ResultsEight patients were enrolled. The mean baseline train-of-four ratio was 1.02 (SD=0.04) for the TwitchView and 0.99 (SD=0.03) for the TetraGraph (p=0.22). Bland Altman analysis of all of the train-of-four ratio data found that average TwitchView train-of-four values were larger with a bias of 0.10. Train-of-four counts and train-of-four ratios were generally less when measured with TetraGraph than when measured with TwitchView.In 83% (209/253) of data pairs, the result from TetraGraph was less than the result from TwitchView and in 6% (16/253) of data pairs, the result from TetraGraph was greater than the result from TwitchView (p<0.0001). In 11% (28/253) of data pairs, the result from TetraGraph was the same as TwitchView [95%CI 7.35% 16.0%]. Evaluation of individual patient results confirmed the overall results. In some cases there were large discrepancies, such as 4 twitches reported by the TwitchView when the TetraGraph reported a post-tetanic count. ConclusionsUsers of the TetraGraph electromyograph should be aware that significant underestimation of post-tetanic-count, train-of-four count and train-of-four ratio may occur. This could result in administration of unnecessary reversal agents, excessive doses of reversal agents, or delay in extubation. We are undertaking a comparison of the TetraGraph monitor to mechanomyography to confirm the results of this pilot study.
Matching journals
The top 3 journals account for 50% of the predicted probability mass.
Similar papers in this journal
- Characteristics of motor evoked potentials in patients with peripheral vascular disease. 93%
- Virtual reality as a strategy for intra-operatory anxiolysis and pharmacological sparing in patients undergoing breast surgeries: the V-RAPS randomized controlled trial protocol 93%
- Recognition and Differentiation of Dural Puncture Click Sensation: A Subjective and Objective Prospective Study of Dural Puncture Forces Using Fine-Gauge Spinal Needles 93%
Similar papers in this journal
- Injury-Prone: Peripheral nerve injuries associated with prone positioning for COVID-19-related acute respiratory distress syndrome 92%
- The impact of female sex on anaesthetic awareness, depth and emergence: A systematic review and meta-analysis 91%
- Effect of Telemedicine Support for Intraoperative Anaesthesia Care on Postoperative Outcomes: The TECTONICS Randomised Clinical Trial 90%
Similar papers in this journal
- A Systematic Investigation of Detectors for Low Signal-to-Noise Ratio EMG Signals 91%
- Cardiorespiratory physiological perturbations after acute smoke-induced lung injury and during extracorporeal membrane oxygenation support in sheep 89%
- Development and Validation of a Bedside Scale for Assessing Upper Limb Function Following Stroke: A Methodological Study 89%
Similar papers in this journal
- Anesthetic agents affect urodynamic parameters and anesthetic depth at doses necessary to facilitate preclinical testing in felines 93%
- Sensory and motor electrophysiological mapping of the cerebellum in humans. 92%
- Pre-movement changes in sensorimotor beta oscillations predict motor adaptation drive. 91%
Similar papers in this journal
- A Predictive Nomogram for In-ICU Deterioration of Stage 1 Pressure Injuries: A Retrospective Study 88%
- Modified Preparation Method of Ideal Platelet-Rich Fibrin Matrix (PRFM) from Whole Blood 88%
- Regional pleural strain measurements during mechanical ventilation using ultrasound elastography: A randomised, crossover, proof of concept physiologic study 88%
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.