Bench Testing of an 8 N Slip Resistance Specification for the Handle Shaft Locking Interface of an Articulated Laparoscopic Instrument
Ma, H.
Show abstract
Background: Articulated laparoscopic instruments require a reliable mechanical connection between the handle and shaft to maintain the intended instrument configuration during surgical manipulation. The handle shaft locking interface should resist unintended relative displacement under external loading. However, the appropriateness of a specific slip resistance threshold must be evaluated in relation to the instrument design and intended conditions of use. Objective: To evaluate the mechanical performance associated with an 8 N slip resistance specification for the locked handle shaft interface of an articulated laparoscopic instrument using a controlled bench model. Methods: A bench based mechanical evaluation method was designed using the instrument under test, mechanical fixtures, a calibrated force measurement device, and displacement measurement equipment. After engagement of the handle shaft locking mechanism, external forces were applied to evaluate resistance to relative displacement. Testing included axial loading, loading in relevant transverse directions, and repeated locking unlocking cycles. The primary outcome was the force corresponding to the onset of relative slippage at the locked interface. Secondary outcomes included displacement under load, complete locking failure, and changes in mechanical performance after repeated operation. Acceptance criteria should be established before confirmatory testing. Results: All 30 instruments from three production batches met the 8 N axial requirement (30/30). The axial slip onset force was 12.40 N . Under a sustained load of 8 N for 60 s, mean relative displacement increased from 0.06 mm to 0.10 mm, and mean residual displacement was 0.02 mm. After 1,000 cycles, the axial slip-onset force was 12.00 N . Conclusions: This study provides a bench-testing framework for evaluating an 8 N slip resistance specification for the handle shaft locking interface of an articulated laparoscopic instrument. Whether this specification provides adequate mechanical stability can be determined only by comparing the measured performance with engineering acceptance criteria established in advance on the basis of the instrument's intended use and foreseeable loading conditions.
Matching journals
The top 4 journals account for 50% of the predicted probability mass.
Similar papers in this journal
Similar papers in this journal
- Precision in Spinal Cord Injury Research: A Novel Electromagnetic Impactor for a Consistent Porcine Model 94%
- Fully Implanted Prostheses for Musculoskeletal Limb Reconstruction after Amputation: An In Vivo Feasibility Study 92%
- Urological benchtop and in silico models validated by human penile tissue inflation tests 90%
Similar papers in this journal
Similar papers in this journal
Similar papers in this journal
- Validity of intraoperative imageless navigation (Naviswiss™) for component positioning accuracy in primary total hip arthroplasty: Protocol for a prospective observational cohort study in a single-surgeon practice 92%
- Comparative Computed Tomography with Stress Manoeuvres for Diagnosing Distal Isolated Tibiofibular Syndesmotic Injury in Acute Ankle Sprain: a Protocol for an Accuracy-Test Prospective Study. 90%
- Doctors of chiropractic working with or within integrated health care delivery systems: a scoping review protocol 90%
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.