Muse 2 forehead photoplethysmography agrees with Polar H10 for heart rate but overestimates ultra-short RMSSD

Griadhi, I. P. A.; Suadnyana, I. A. A.; Wahyuni, N.

2026-10-08 sports medicine
10.64898/2026.10.05.26364782 medRxiv
Show abstract

The Muse 2 EEG headset also records forehead photoplethysmography (PPG). We tested whether its pulse intervals can replace Polar H10 RR intervals for ultra-short heart-rate variability in 28 seated recordings (60 s eyes open, then 60 s eyes closed), using fixed processing, no exclusions and eleven sensitivity analyses. Mean heart rate agreed almost perfectly (ICC [≥] 0.999; bias -0.22 and 0.00 beats min-1). SDNN agreement was good to excellent (ICC 0.86 and 0.91; bias +8.6 and +8.2 ms). RMSSD agreement was moderate (ICC 0.60): Muse exceeded Polar in at least 27 of 28 sessions (bias +19.2 and +18.5 ms; about 1.9-fold), and the ratio bias fell as RMSSD rose. Interval differences (SD 23 ms) implied about 16 ms of independent timing error per pulse, which reproduced the RMSSD bias; without ECG, sensor jitter could not be separated from variation in pulse transit time. Tracking of the eyes-closed change in ln RMSSD was not demonstrated ({rho} = 0.35, 95% CI -0.07 to 0.67; detectable {rho} [≥] 0.51). Muse 2 PPG can supply mean heart rate and, with a stated bias, SDNN; its ultra-short RMSSD is not interchangeable with Polar RMSSD and should be reported as pulse-rate variability.

Matching journals

The top 3 journals account for 50% of the predicted probability mass.

50% of probability mass above

"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.