Surgically Modifiable Insertion Geometry Drives Thrombogenic Flow in the Modified Blalock-Taussig-Thomas Shunt

Qiao, Y.; Penn, E.; Miller, J.; Bugenhagen, S.; Rohatgi, R.; Mercer, K.; Kulp, B.; Eghtesady, P.; Genin, G. M.; Bark, D. L.; Rabinowitz, E.

2026-09-15 cardiovascular medicine
10.64898/2026.09.11.26362753 medRxiv
Show abstract

ABSTRACT Background: The modified Blalock-Taussig-Thomas shunt (mBTTS) sustains pulmonary blood flow in infants with cyanotic or single-ventricle heart disease, but shunt thrombosis occurs in 8-12% of cases and carries substantial mortality. Systemic anticoagulation has limited efficacy. Objectives: This study tested whether surgically modifiable shunt geometry might form a second, patient-specific approach to maintaining shunt patency, via management of thrombogenic hemodynamics. Methods: Three-dimensional mBTTS anatomies were reconstructed from computed tomography in 10 infants (4 thrombosed, 6 patent). Pulsatile inlet waveforms derived from Doppler ultrasound were coupled with three-element Windkessel outlet models individually calibrated to catheter-derived pressures. Transient computational fluid dynamics simulations predicted wall shear rate (WSR) and elongational strain rate (ESR). Associations between surgically relevant geometric parameters and hemodynamic metrics were evaluated. Results: Simulated pressures closely matched clinical measurements at all four outlets in all 10 models. Despite substantial anatomic variability, peak WSR and ESR consistently localized to the shunt-subclavian junction in 8 of 10 patients, including all 4 thrombosed shunts. Greater deviation of shunt insertion angle from perpendicular was associated with higher systolic and cycle-averaged normalized WSR ({rho} = 0.94, p < 0.001; {rho} = 0.88, p = 0.002) and ESR ({rho} = 0.82, p = 0.007; {rho} = 0.70, p = 0.03). Larger shunt diameter was associated with lower WSR and ESR. Thrombosed shunts demonstrated higher shear-related metrics than patent shunts, particularly during systole. Conclusions: In patient-specific mBTTS anatomies, insertion angle and shunt diameter are determinants of local abnormal flow, concentrated at the shunt-subclavian junction. Patient-specific hemodynamic assessment may inform shunt construction and interstage risk stratification.

Matching journals

The top 5 journals account for 50% of the predicted probability mass.

50% of probability mass above

"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.