Readout-dependent time-of-flight weighting of absorption and flow in interferometric diffuse optics
Marzejon, M.; Mogharari, N.; Nowacka-Pieszak, K.; Borycki, D.
Show abstract
Joint optical measurement of hemoglobin and blood-flow dynamics could provide a compact route to richer functional monitoring of tissue physiology. However, hybrid diffuse-optical systems commonly optimize photon time-of-flight (TOF) for individual readouts, even when absorption and flow are recovered from the same acquisition. Here, we asked whether a single TOF window can optimally preserve both signals. Dual-wavelength TOF-iSCOS recovered absorption, hemoglobin and relative blood-flow index (BFI) from the same interferometrically sensed optical field, while Early, Middle and Late photon windows provided progressively greater weighting of longer and deeper photon paths. During forearm cuff occlusion in 11 adults, absorption and flow showed markedly different TOF dependence: Late-gate BFI responses were retained, whereas absorption decreased to a median Late/Early ratio of 0.18. The resulting readout-by-TOF interaction was positive in all 11 participants (median 2.61 octaves, 95% bootstrap CI 2.44-3.22; exact sign-test P = 9.8 x 10-4). Readout-specific gate selection increased held-out response retention by 24.8% (95% CI 8.8-42.5%) relative to a single common gate. We then tested the approach during prefrontal working-memory activation in seven participants using a higher-load 2-back task relative to a low-load 0-back control. Late-gate BFI increased in all seven participants (+5.07%, 95% CI 1.46-8.68%; P = 0.014) and remained positive after adjustment using an Early-TOF signal, whereas hemoglobin responses were less consistent. Layered Monte Carlo modeling showed that the same detected photon histories acquire different absorption and dynamic-scattering sensitivity across TOF, linking the experimental effects to anatomy- and geometry-conditioned tissue weighting. Within the present geometry and photon budget, the central result is therefore not that one readout is intrinsically superior, but that absorption and flow recovered from the same optical field retain physiological information differently across photon time-of-flight.
Matching journals
The top 5 journals account for 50% of the predicted probability mass.
Similar papers in this journal
- Time-Domain Diffuse Optical Tomography for Precision Neuroscience 94%
- Functional ultrasound imaging and neuronal activity: how accurate is the spatiotemporal match? 89%
- Laminar multi-contrast fMRI at 7T allows differentiation of neuronal excitation and inhibition underlying positive and negative BOLD responses 89%
Similar papers in this journal
- Multimodal retinal imaging by visible light optical coherence tomography and phosphorescence lifetime ophthalmoscopy in the mouse eye 92%
- Widefield in vivo imaging system with two fluorescence and two reflectance channels, a single sCMOS detector, and shielded illumination 91%
- Characterization of red fluorescent reporters for dual-color in vivo three-photon microscopy 91%
Similar papers in this journal
- Characterization of a fiber-coupled SPAD camera system for deep-tissue blood-flow measurement using diffuse correlation spectroscopy 94%
- A quasi-analytic solution for real-time multi-exposure speckle imaging of tissue perfusion 92%
- High-speed adaptive optics line-scan OCT for cellular-resolution optoretinography 91%
Similar papers in this journal
- Kidney Tissue Characterization using Normalized Raman Imaging and Segment-Anything 92%
- Wide and Deep Imaging of Neuronal Activities by a Wearable NeuroImager Reveals Premotor Activity in the Whole Motor Cortex 91%
- Scalable cytoarchitectonic characterization of large intact human neocortex samples 91%
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.