cFAR and Relative Signal: Diagnosing Preferred Orientation in Single-Particle Cryo-EM
Peretroukhin, V.; McLean, M.; Punjani, A.
Show abstract
The quality of single particle cryo-EM reconstructions can be severely degraded when an insufficient variety of 3D particle orientations is present in the image data, limiting downstream model building and interpretation. However, it is often difficult to ascertain whether or not a particular dataset suffers from such preferred orientation since the required orientation coverage depends on target geometry, alignment accuracy, and particle quality. To simplify diagnosis of preferred orientation, we present two complementary methods. First, the conical Fourier Shell Correlation Area Ratio (cFAR) compares the worst- and best-correlating conical regions of 3D Fourier space to quantify half-map anisotropy into a single, easily interpretable score ranging from zero to one. Second, Relative Signal, a companion to cFAR, directly relates signal content to viewing direction so that under-sampled views can be identified. We characterize our methods and compare them to existing anisotropy detection approaches on synthetic data and on 14 real datasets that span sundry molecular weights and structure types. Implementations of both cFAR and Relative Signal are included in CryoSPARC v4.5 and later versions.