Comparative Thermal Effects of Single Shot Pulsed Field Ablation Systems using a Thermochromic Hydrogel
Gill, J.; Saija, C.; Sagar, V.; Zuberi, Z.; Bajpai, A.; Rhode, K.; Leung, L. W.; Gallagher, M. M.
Show abstract
Background Pulse-field ablation (PFA) is regarded as a non-thermal ablation modality, but there is an increasing range of complications that could be due to thermal effects. Methods The hydrogel undergoes permanent colour change when a target temperature is reached allowing direct visualisation of the surface thermal footprint and depth. Comparative lesion sets using a variable loop circular catheter (VP), circular over-the-wire catheter (PS) and pentaspline catheter (FP) were performed. Protocols included single and stacked applications with variation of force, irrigation, and voltage. The hydrogel lesions were analysed en-face and by section using digital image analysis. Results All 3 PFA catheters tested had significant thermal footprints. The VP catheter had the largest mean surface footprint (156.1mm2) and thermal depth (1.31mm) compared to the other two catheters (PS 55.4mm2 & 1.1mm, FP 29.8mm2 & 1.05mm, p<0.005). Increasing irrigation showed a trend to reduce thermal footprint but did not achieve statistical significance. Increasing voltage increased thermal footprint, but increasing force had negligible effect. Stacked lesions incrementally increased thermal lesion footprint and depth in all catheters. Thermal depths of up to 2.4mm were observed. Areas of darkening and degradation of the hydrogel were observed with the VP and FP catheters, consisting of up to 47% of lesion area. No darkening was observed with the PS catheter. Conclusions There are significant thermal footprints in all the systems tested. Temperatures exceeding 60oC have been demonstrated, comparable to radiofrequency ablation, and this may explain the mechanism of injury in some reports of collateral damage during PFA.
Matching journals
The top 6 journals account for 50% of the predicted probability mass.
Similar papers in this journal
- Reliability and Performance of the IRRAflow® System for Intracranial Lavage and Evacuation of Hematomas - A Technical Note 94%
- Surface Anticoagulation of Mechanical Heart Valves using Electrically Induced Biomimetic Glycocalyx: an In-vitro study to assess Hemocompatibility and Optimal Voltage 92%
- Flat Electrode Contacts for Peripheral Nerve Stimulation 92%
Similar papers in this journal
- A Novel Temperature Controlled Bipolar Radiofrequency Ablation: An Ex Vivo Study for Optimizing Efficacy and Safety Parameters 96%
- Impact of transseptal puncture location on the fossa ovalis on first-pass pulmonary vein isolation 92%
- Procedural Volume and Outcomes with Atrial Fibrillation Ablation: A Report from the NCDR AFib Ablation Registry 91%
Similar papers in this journal
- Patient Screening for Self-Expanding Percutaneous Pulmonary Valves using Virtual Reality 92%
- Non-Invasive Scale Measurement of Cardiac Output Compared with the Gold-Standard Direct Fick Method: A Feasibility Study 91%
- Smartwatch Facilitated Remote Health Care for Patients Undergoing Transcatheter Aortic Valve Replacement Amid COVID-19 Pandemic 90%
Similar papers in this journal
- Endoluminal catheter pulsed field ablation for the treatment of atherosclerotic vascular disease 92%
- Switching the left and the right hearts: A novel bi-ventricle mechanical support strategy with spared native single-ventricle 91%
- Urological benchtop and in silico models validated by human penile tissue inflation tests 91%
Similar papers in this journal
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.