Comparison of Mechanical Tissue Properties Using MyotonPRO and Time-Harmonic Elastography: Understanding Fundamental Differences and Statistical Relationships
Kurz, E.; Valli, G.; Meyer, T.; Proger, S.; Schwesig, R.; Bartels, T.; Delank, K.-S.; Sack, I.; Aghamiry, H. S.
Show abstract
Abstract Purpose: MyotonPRO (MTP) and time-harmonic elastography (THE) are increasingly used to assess muscle mechanical properties, yet they operate on fundamentally different physical principles. MTP measures composite MTP stiffness (N/m) through surface oscillations, while THE quantifies intrinsic shear modulus (THE stiffness, kPa) via propagating shear waves. This study aimed at systematically compare MTP and THE measurements in the vastus lateralis muscle across different contraction intensities and examine how the skin layer and subcutaneous fat (SLSF) thickness influence their relationship. Methods: Twenty-six healthy adults (15 males, 11 females; age 25 [SD 4] years) underwent MTP and THE measurements of the vastus lateralis at rest and during isometric contractions at 15% and 30% maximal voluntary contraction (MVC). Effects of contraction intensities on tissue properties were assessed using univariate analyses of variance with repeated measures. Associations between the different outcomes of THE and MTP technologies were explored using Pearson's correlations and partial correlation coefficients separately for each contraction intensity with adjustment of the SLSF thickness of participants. Results: Both technologies detected contraction intensity-dependent stiffening across all outcomes (p < 0.001). THE stiffness increased from 5.3 [1.2] kPa at rest to 15.6 [6.1] kPa at 30% MVC; THE wave attenuation increased from 0.83 [0.19] to 1.42 [0.36] s/m while MTP stiffness increased from 337.3 [49.3] N/m at rest to 529.4 [160.7] N/m at 30% MVC. Correlations between modalities were weak and condition-dependent. THE wave attenuation did not significantly correlate with any MTP outcome across conditions. Conclusion: MTP and THE detect contraction-induced stiffening through fundamentally different physical mechanisms and should not be regarded as interchangeable. Their correlation is modest at rest and breaks down (or reverses) during active contraction, with subcutaneous fat as a key modifying factor. Clinical trial number: Not applicable.
Matching journals
The top 2 journals account for 50% of the predicted probability mass.