Scan length as a major driver of CT radiation dose: a diagnostic reference level audit from Kosovo

Rudi, G.; Vula, F.; Bicaku, A.; Dedushi, K.; Ahmetgjekaj, I.

2026-05-17 radiology and imaging
10.64898/2026.05.12.26353024 medRxiv
Show abstract

Computed tomography is the largest contributor to population radiation dose from medical imaging, yet no diagnostic reference levels (DRLs) have been published from Kosovo or the Western Balkans. This retrospective audit analyzed all CT examinations performed on a 128- slice scanner at the University Clinical Centre of Kosovo between January and March 2026. After exclusions, 1,535 acquisitions from 1,092 patients across nine examination categories were analyzed. Local DRLs were defined as the 75th percentile and compared against German (BfS 2022) and Turkish (Kahraman et al., 2024) reference values. Head CT (n = 590) demonstrated CTDIvol 4.7% below the BfS DRL yet scan length 98.5% above the orientation value (median 25.8 vs 13 cm). Abdomen-pelvis CTDIvol matched the BfS reference while scan length exceeded it by 28%. Coronary CTA showed CTDIvol +377%, consistent with retrospective ECG gating. Excess scan length, not CTDIvol, is the major driver of elevated dose at this institution. The identified excesses are correctable through technologist landmarking training, protocol review, and enabling iterative reconstruction.

Published in Radioprotection · not in our set (fewer than 10 published preprints to learn from) · training set

Matching journals

The top 5 journals account for 50% of the predicted probability mass.

50% of probability mass above

"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.