An analysis of retinal safety when using a laser based low-level red light therapy device for myopia
Schulmeister, K.; Marshall, J.
Show abstract
PurposeTo evaluate the retinal safety of repeated low-level red-light (RLRL) therapy using the Eyerising Myopia Management Device (EMMD) by analysing exposure parameters relative to established thermal and photochemical retinal injury thresholds and empirical human exposure data. MethodsEmission characteristics of the EMMD were measured in an accredited laboratory under worst-case conditions. Parameters assessed included wavelength, intraocular power, corneal irradiance, and retinal image characteristics across accommodative states. These measurements were compared with international safety standards, maximum permissible exposure limits, and experimentally derived retinal injury thresholds from animal studies and validated computational models. The effects of repeated exposures from RLRL therapy using the EMMD were evaluated using photochemical additivity principles and repair kinetics, and further contextualised using human volunteer exposure data. ResultsThe EMMD emitted red laser radiation at 654-655 nm with a maximum intraocular power of approximately 1 mW through a 7 mm pupil, placing it within Class 3R and marginally above the Class 2 limit. Corneal irradiance was approximately 26 W m- 2, well below conservative photochemical exposure limits. Thermal injury modelling indicated retinal damage thresholds above device exposure, including under worst-case assumptions of minimal retinal image size and absence of eye movements. Accounting for repeated daily exposures and photochemical additivity, safety margins remained approximately 3-fold for a 7 mm pupil and approximately 8-fold for a more realistic 4 mm pupil. Human volunteer studies demonstrated no detectable structural or functional retinal injury at exposure levels approximately five times higher than those produced by the EMMD. ConclusionExposure parameters of RLRL therapy using the EMMD remain well below conservative retinal injury thresholds under prescribed use conditions. Integration of experimental, modelling, and human data indicates substantial safety margins, supporting its safe clinical use.
Matching journals
The top 1 journal accounts for 50% of the predicted probability mass.
Similar papers in this journal
- Ocular and Facial Far-UVC Doses from Ceiling-Mounted 222 nm Far-UVC Fixtures 97%
- 222 nm Far-UVC from filtered Krypton-Chloride excimer lamps does not cause eye irritation when deployed in a simulated office environment 96%
- Wavelength-dependent DNA photodamage in a 3-D human skin model over the far-UVC and germicidal-UVC wavelength ranges from 215 to 255 nm 96%
Similar papers in this journal
Similar papers in this journal
- Cell viability assessment associated with a contact of gas bubbles produced by femtosecond laser breakdown in cell culture media 95%
- The virucidal effects of 405 nm visible light on SARS-CoV-2 and influenza A virus 94%
- Scalable, effective, and rapid decontamination of SARS-CoV-2 contaminated N95 respirators using germicidal ultra-violet C (UVC) irradiation device 94%
Similar papers in this journal
- Two-Photon Microperimetry: A media opacity-independent retinal function assay 92%
- Visual field evaluation using Zippy Adaptive Threshold Algorithm (ZATA) Standard and ZATA Fast in patients with glaucoma and healthy individuals 92%
- Evaluating human rod photoreceptor function using pixelwise intensity-based optoretinography 91%
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.