Computational and Experimental Evaluation of a Flow-Conditioning Anastomotic Device for Arteriovenous Fistula Maturation
Saum, K. L.; Campos, B.; Celdran-Bonafonte, D.; Oren, L.; Owens, A. P.; Roy-Chaudhury, P.
Show abstract
The arteriovenous fistula (AVF) is the preferred method of vascular access for hemodialysis; however, 30-50% of AVFs undergo primary failure and are unsuitable for clinical use. As disturbed hemodynamics initiate endothelial injury and intimal hyperplasia, we designed an endovascular flow-conditioning anastomotic device (FCAD) to directly improve AVF hemodynamics and protect the anastomotic region. Using computational fluid dynamics, we characterized the flow field and wall shear stress (WSS) profiles in idealized AVF models with and without the FCAD. Incorporation of the FCAD into a brachiocephalic AVF model reduced regions of oscillatory WSS and generated a symmetrical flow profile in the draining vein compared to a reference AVF. Parametric studies also identified an FCAD geometry with a tab angle, height, and aspect ratio of 30{degrees}, 0.1 diameters, and 1.0 restored time-averaged WSS along the inner venous wall, achieving a physiological level without inducing regions of oscillatory flow throughout the cardiac cycle. Similar findings were observed with an in vitro model using particle imaging velocimetry. This study demonstrates the feasibility of the FCAD to normalize venous flow and WSS while imposing minimal resistance to blood flow. Restoring physiological WSS levels on the venous wall is expected to preserve endothelial function and improve AVF maturation.
Matching journals