Systematic Errors in DC Offset Measurement and Mitigation at Neurostimulation Electrodes: Causes, Implications and Solutions

Gholston, A. K.; Sanchez, R. A.; Bigger, T. B.; Averbeck, S. R.; Rieth, L.; Ludwig, K. A.; Trevathan, J. K.

2025-12-11 neuroscience
10.64898/2025.12.08.691908 bioRxiv
Show abstract

ObjectiveNeuromodulation therapies are becoming increasingly common as a therapeutic strategy to treat neurological disorders and deficits. Despite their widespread use, the frameworks used to evaluate device longevity and electrochemical safety are limited, producing variable results for platinum (Pt) electrodes across studies. In addition, next-generation neuromodulation electrodes like thin-film microelectrode arrays (TFMEAs) are more prone to failure by virtue of their smaller size, higher edge to bulk ratios, and greater sensitivity to electrochemical dissolution processes. Two potential contributors to the variability in reported electrochemical safety limits and device performance are particularly sensitive to scaling electrodes to smaller geometries, namely: 1) insufficient measurement device input impedance and 2) unmitigated stimulator leakage currents. ApproachTo explore these effects, we first characterize the impact of electrode size (1.27 x 10-4 to 7.85 x 10-3 cm2) on apparent open circuit potential (VaOCP) and interpulse potential (VIPP) measurements at Pt electrodes across measurement instrumentation configurations. Next, we evaluated the effectiveness of commonly used DC mitigation strategies, including stimulation with capacitive coupling (CC), and with capacitive coupling combined with cathode/anode shorting after each pulse (CC+ES), as a function of electrode size. Main ResultsWithout proper DC mitigation, insufficient measurement input impedance leads to an underestimation of DC offsets ({Delta}VIPP) and, consequently, overestimation of charge injection limit (Qinj) at Pt electrodes. In contrast, when DC offsets are properly mitigated and the input impedance of the measurement device is sufficient, the experimentally derived charge injection limits for Pt electrodes are larger than what has been reported historically. SignificanceBy identifying the instrumentation-related confounds biasing measurements of electrochemical safety limits, this work provides a more consistent basis for the assessment of Qinj for safety, as well as a framework for identifying and mitigating DC offset as a potential source of variability in electrophysiological findings across studies using TFMEAs.

Matching journals

The top 1 journal accounts for 50% of the predicted probability mass.

50% of probability mass above

"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.