Back

In Vivo Test-retest Quantitative Characterization of Echo Planar Imaging Apparent Diffusion Coefficient Reproducibility for Head and Neck Cancers on a 1.5T MR-Linac Platform: Technical Validation using QIBA Metrology

2025-03-07 radiology and imaging Title + abstract only
View on medRxiv
Show abstract

Background and PurposeTo detect changes in apparent diffusion coefficient (ADC) values during radiation therapy for biological image-guided adaptive radiation therapy, the variability in ADC must be characterized. We evaluated the reproducibility of ADC values in head and neck cancers on a 1.5T MR-linac. Methods39 head and neck cancer patients (36 primary tumors, 55 lymph nodes) were imaged with echo-planar imaging diffusion-weighted MRI on a 1.5T MR-linac at two time points before the start of...

Predicted journal destinations