Optimising Cardiac Diffusion Tensor Imaging In Vivo: More Directions or Averages?
Coveney, S.; Shelley, D.; Foster, R. J.; Afzali, M.; Poenar, A.-M.; Sharrack, N.; Plein, S.; Dall'Armellina, E.; Schneider, J. E.; Nguyen, C.; Teh, I.
Show abstract
BackgroundCardiac diffusion tensor imaging (cDTI) is sensitive to imaging parameters including the number of unique diffusion encoding directions (ND) and number of repetitions (NR; analogous to number of signal averages or NSA). However, there is no clear guidance for optimising these parameters in the clinical setting. MethodsSpin echo cDTI data with 2nd order motion compensated diffusion encoding gradients were acquired in ten healthy volunteers on a 3T MRI scanner with different diffusion encoding schemes in pseudo-randomised order. The data were subsampled to yield 96 acquisition schemes with 6 [≤] ND [≤] 30 and 33 [≤] total number of acquisitions (NAall) [≤] 180. Stratified bootstrapping with robust fitting was performed to assess the precision and accuracy of each acquisition scheme. This was quantified across a mid-ventricular short-axis slice in terms of root mean squared difference (RMSD) with respect to the full reference dataset, and standard deviation (SD) across bootstrap samples respectively. ResultsFor the same acquisition time, the ND = 30 schemes had on average 48%, 40%, 34% and 34% lower RMSD and 6.2%, 7.4%, 10% and 5.6% lower SD in MD, FA, HA and E2A compared to the ND = 6 schemes. Given a fixed number of high b-value acquisitions, there was a trend towards lower RMSD and SD of MD and FA with increasing numbers of low b-value acquisitions. Higher NAall with longer acquisition times led to improved accuracy in all metrics whereby quadrupling NAall from 40 to 160 volumes led to a 21%, 40%, 13% and 4.5% reduction in RMSD of MD, FA, HA and E2A, averaged across six diffusion encoding schemes. Precision was also improved with a corresponding 53%, 50%, 52% and 36% reduction in SD. ConclusionsWe observed that accuracy and precision were enhanced by (i) prioritising number of diffusion encoding directions over number of repetitions given a fixed acquisition time, (ii) acquiring sufficient low b-value data, (iii) using longer protocols where feasible. For clinically relevant protocols, our findings support the use of ND = 30 and NAb50:NAb500 [≥] 1/3 for better accuracy and precision in cDTI parameters. These findings are intended to help guide protocol optimisation for harmonisation of cDTI.
Matching journals
The top 2 journals account for 50% of the predicted probability mass.
Similar papers in this journal
- Combined Angiographic, Structural and Perfusion Radial Imaging using Arterial Spin Labeling 97%
- The Effect of Oblique Image Acquisition on the Accuracy of Quantitative Susceptibility Mapping and a Robust Tilt Correction Method 97%
- A characterization of cardiac-induced noise in R2* maps of the brain 97%
Similar papers in this journal
- Non-negative least squares computation for in vivo myelin mapping using simulated multi-echo spin-echo T2 decay data 95%
- Multiple sclerosis cortical lesion detection with deep learning at ultra-high-field MRI 95%
- Hierarchical Bayesian Modelling Improves Microstructural Parameter Mapping in Diffusion and Exchange MRI Data 95%
Similar papers in this journal
- Deep learning microstructure estimation of developing brains from diffusion MRI: a newborn and fetal study 96%
- SpinFlowSim: a blood flow simulation framework for histology-informed diffusion MRI microvasculature mapping in cancer 94%
- dStripe: slice artefact correction in diffusion MRI via constrained neural network 94%
Similar papers in this journal
Similar papers in this journal
- Calgary Normative Study: Study Design Of A Prospective Longitudinal Study To Characterize Potential Quantitative Mr Biomarkers Over The Adult Lifespan 89%
- Response of the Myocardium to Hypertrophic Conditions in the Adult Population: The Prospective Observational REMODEL Cohort Study Protocol 89%
- Early Structural Cardiovascular Disease, HIV, and Tuberculosis in East Africa (ASANTE): Cross-sectional study protocol for a multimodal cardiac imaging study in Nairobi, Kenya 85%
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.