A Method for Electrical Stimulus Artifact Removal Exploiting Neural Refractoriness: Validation by Contrasting Cathodic and Anodic Stimulation
Nakhmani, A.; Block, J.; Awad, M.; Olson, J.; Smith, R. J.; Bentley, N. J.; Holland, M. T.; Brinkerhoff, S. A.; Gonzalez, C. L.; Moffitt, M.; Walker, H. C.
Show abstract
ObjectiveTo present a novel method for removing stimulus transient that exploits the absolute refractory period of electrically excitable neural tissues. BackgroundElectrical stimulation often generates significant signal artifacts that can obscure important physiological signals. Removal of the artifact and understanding latent information from these signals could provide objective measures of circuit engagement, potentially driving advancements in neuromodulation research and therapies. MethodsWe conducted intracranial physiology studies on five consecutive patients with Parkinsons disease who underwent deep brain stimulation (DBS) surgery as part of their routine care. Monopolar stimuli (either cathodic or anodic) were delivered in pairs through the DBS electrode across a range of inter-stimulus intervals. Recordings from adjacent unused electrode contacts used broadband sampling and precise synchronization to generate a robust template for the stimulus transient during the absolute refractory period. These templates of stimulus transient were then subtracted from recordings at different intervals to extract and analyze the residual neural potentials. ResultsAfter artifact removal, the residual signals exhibited absolute and relative refractory periods with timing indicative of neural activity. Cathodic and anodic DBS pulses generated distinct patterns of local tissue activation, showing phase independence from the prior stimulus. The earliest detectable neural responses occurred at short peak latencies (ranging from 0.19 to 0.38 ms post-stimulus) and were completely or partially obscured by the stimulus artifact prior to removal. Cathodic stimuli produced stronger local tissue responses than anodic stimuli, aligning with clinical observations of lower activation thresholds for cathodic stimulation. However, cathodic and anodic pulses induced artifact patterns that were equivalent but opposite. InterpretationThe proposed artifact removal technique enhances prior approaches by allowing direct measurement of local tissue responses without requirements for stimulus polarity reversal, template scaling, or specialized filters. This approach could be integrated into future neuromodulation systems to visualize stimulus-evoked neural potentials that would otherwise be obscured by stimulus artifacts.
Matching journals
The top 2 journals account for 50% of the predicted probability mass.
Similar papers in this journal
- Controlling pallidal oscillations in real-time in Parkinson's disease using evoked interference deep brain stimulation (eiDBS): proof of concept in the human 97%
- Real-time suppression and amplification of frequency-specific neural activity using stimulation evoked oscillations 97%
- Automated artifact injection into sensing-capable brain modulation devices for neural-behavioral synchronization and the influence of device state 96%
Similar papers in this journal
- Towards adaptive deep brain stimulation: clinical and technical notes on a novel commercial device for chronic brain sensing 96%
- Automated optimization of deep brain stimulation parameters for modulating neuroimaging-based targets 95%
- Microneurography as a Minimally Invasive Method to Assess Target Engagement During Neuromodulation 95%
Similar papers in this journal
- Closed-loop control of functional electrical stimulation using a selectively recording and bidirectional nerve cuff interface 92%
- Forecast Padding Enhances Accuracy and Robustness of EEG-Phase-Synchronized TMS 92%
- A BCI-based vibrotactile neurofeedback training improves motor cortical excitability during motor imagery 91%
Similar papers in this journal
- Endogenous signals during active movement predict deep brain stimulation evoked potential pathways: Results of a transfer function analysis 97%
- Optimal Multichannel Artifact Prediction and Removal for Brain Machine Interfaces and Neural Prosthetics 95%
- Similarities Between Somatosensory Cortical Responses Induced via Natural Touch and Microstimulation in the Ventral Posterior Lateral Thalamus in Macaques 93%
Similar papers in this journal
- Proper reference selection and re-referencing to mitigate bias in single pulse electrical stimulation data 95%
- A biophysically constrained brain connectivity model based on stimulation-evoked potentials. 94%
- Stimulation-Evoked Effective Connectivity (SEEC): An in-vivo approach for defining mesoscale corticocortical connectivity 94%
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.