Reliability and criterion validity of a low-cost handgrip dynamometer: The Camry
Sanchez-Aranda, L.; Fernandez-Ortega, J.; Martin-Fuentes, I.; Jurak, G.; Ruiz, J. R.; Csanyi, T.; Ortega, F. B.
Show abstract
BackgroundHandgrip strength has been related with multiple health outcomes, including all-cause mortality and morbidity. Handgrip testing is a highly valid and reliable method, included in evidence-based fitness test batteries from preschool to older ages. Previously, Jamar and TKK dynamometers have shown good reliability and validity against known weights. However, the cost of these dynamometers is the major limitation for implementing handgrip strength testing in certain countries and settings, as well as at large scale. Recently, a ten times cheaper model (Camry Dynamometer) has been used in fitness surveillance systems, though its reliability and validity, compared to known weights and other well-validated dynamometers, remains unknown. Therefore, the aims of the current study were to test to examine test-retest reliability, inter-model reliability (comparing a Camry dynamometer with 3000 uses versus a new Camry dynamometer), and inter-instrument reliability (Camry versus TKK dynamometer) of Camry dynamometer, using calibrated known weights. MethodsA digital TKK 5401 dynamometer and two Camry EH101, a new and an "old" (3000 uses), dynamometers were used. Intra-instrument and inter-instrument reliability, and criterion related validity were assessed comparing the measures of the dynamometers with calibrated weights using the Bland and Altmans method. ResultsIntra-instrument (retest minus test) reliability was very high (systematic error for test-retest reliability: New Camry = 0.01{+/-}0.49kg; Old Camry = -0.10{+/-}0.49kg; TKK = 0.14{+/-}0.77kg). The comparison between instruments showed small mean differences between Camry dynamometers and TKK (New Camry VS TKK: 0.84{+/-}0.79kg; Old Camry VS TKK: 0.88 {+/-}0.85kg). The systematic error between New and Old Camry dynamometers was 0.03{+/-}0.57kg. Criterion-related validity showed smaller magnitude systematic errors in the Camry than TKK instruments (New Camry: -0.21{+/-}0.35kg; Old Camry: -0.18{+/-}0.79kg; TKK: -1.07{+/-}0.75kg). ConclusionsOur results support that Camry dynamometer has an excellent reliability and validity and is therefore a more affordable alternative for handgrip strength assessment. Our results also indicate a good durability of these devices, and that calibration is not necessary, even after several thousands of uses.
Matching journals
The top 3 journals account for 50% of the predicted probability mass.
Similar papers in this journal
- Are isokinetic leg torques and kick velocity reliable predictors for competitive success in taekwondo athletes? 97%
- Estimation equation of limb lean soft tissue mass in Asian athletes using bioelectrical impedance analysis 97%
- The test-retest reliability and agreement between a fixed frame and belt-stabilised handheld dynamometer for isometric hip flexion and extension peak force measurement in recreational cyclists 96%
Similar papers in this journal
- Muscle activity of cutting manoeuvres and soccer inside passing suggests an increased groin injury risk during these movements 96%
- Joint contact forces during barefoot, minimal and conventional shod running are highly individual 95%
- Association of Muscle Strength to Body Composition Measures using DXA, D 3 Cr, and BIA in Collegiate Athletes 95%
Similar papers in this journal
- The effects of pressure on anticipatory postural adjustments during a jump shot in basketball 96%
- Synergies analysis produces consistent results between motion analysis laboratories 95%
- Differences in centre of mass measurements between markerless and marker-based motion capture systems during balance and mobility assessments in individuals with chronic and sub-acute stroke 95%
Similar papers in this journal
- Translation, adaptation, and measurement properties of the Muscle-Strengthening Exercise Questionnaire among university students in Indonesia 95%
- Reference values and validation of the 1-min sit-to-stand test in healthy 5- to 16-year-old youth: a cross-sectional study 95%
- Protocol for a prospective cohort study on the feasibility of application of nutritional ultrasound in the diagnosis and follow-up of patients with nutritional risk at hospital discharge: Study on body composition and function (DRECO) 95%
Similar papers in this journal
- Reliability of force plate-based measures of standing balance in the sub-acute stage of post-stroke recovery 95%
- Lower complexity of motor primitives ensures robust control of high-speed human locomotion 93%
- Effect of xenon and argon inhalation on erythropoiesis and steroidogenesis: a systematic review 92%
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.