Semi-automated vs. manual: Comparative study of cell culture counting methods using validation parameters

Ramm, S.; Odefey, U.; Frahm, B.; Pein-Hackelbusch, M.

2024-05-31 bioengineering
10.1101/2024.05.30.596619 bioRxiv
Show abstract

Determining cell density and cell viability is fundamental for any cell cultivation process. In addition to the manual counting method using hemocytometers, (semi-)automated methods offer advantages such as lower variability and shortened analysis times. However, these methods should provide at least comparable results to the manual method, which is why a comparison of methods is essential. We conducted a dilution series experimental design according to ISO 20391-2:2019 and compared two cell counting methods based on validation parameters aligned with the ICH Q2(R1) guideline. Regarding specificity and linearity, the manual (hemocytometer) and semi-automated (Countstar BioTech(R)) method exhibited similar results in the two evaluated characteristics total cell density and cell viability of CHO-K1 cells. Regarding repeatability of determining total cell density, the semi-automated method achieved significant ( = 0.05) better results with average relative standard deviations of < 6 %, than the manual method with average relative standard deviations of > 9 %. Concerning repeatability of the cell viability measurement, no significant difference between the two methods were shown. These results show the suitabililty of the dilution series experimental design. For the applied example, they indicate that the investigated semi-automated method is an appropriate alternative to the manual method.

Matching journals

The top 5 journals account for 50% of the predicted probability mass.

50% of probability mass above

"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.