Back

Obtaining high-resolution cryo-EM structures using a common LaB6, 120-keV electron microscope equipped with a sub 200-keV optimised direct electron detector.

Venugopal, H.; Mobbs, J.; Taveneau, C.; Fox, D. R.; Vuckovic, Z.; Knott, G.; Grinter, R.; Thal, D. M.; Mick, S.; Czarnik, C.; Ramm, G.

2024-05-27 biophysics
10.1101/2024.05.26.595910 bioRxiv
Show abstract

Cryo-electron microscopy (Cryo-EM) single particle analysis (SPA) has become a major structural biology technique in recent years. High-resolution cryo-EM typically requires higher voltage cryo-TEMs with coherent FEG sources, stable columns, autoloader systems and direct electron detectors. These setups are specialised for Cryo-EM work and are expensive to establish and maintain. More recently the concept of using 100-keV cryo-TEMs has been introduced as a way to make cryo-EM more affordable and hence accessible to a larger group of researchers. So far, the implementation of these 100-keV cryo-TEMs have relied on specialised microscopes with FEG sources as well as more stable optics than usually present on the common 120-keV TEMs. We here explored whether a standard 120-keV TEM, commonly available at many laboratories worldwide, can be upgraded with a direct electron detector and its suitability for high-resolution cryo-EM using a standard side entry cryo-holder. Using this imaging configuration, we were successful in achieving a 2.65[A] reconstruction for standard apoferritin. We were also able to resolve a more challenging small 64kDa protein haemoglobin to 4.33[A]. Furthermore, we were able to solve an asymmetric 153 kDa membrane protein GPCR (M4 muscarinic acid receptor) to a resolution of 4.4[A]. Importantly, all these results were achieved using a standard automated data collection routine implemented through SerialEM, making it feasible to collect large cryo-EM data sets with a side entry cryo-holder. These results showcase a potentially widely accessible solution to obtaining interpretable cryo-EM structures. Furthermore, we envisage that this imaging configuration gives an option for many EM facilities and laboratories to set up a high-quality cryo-EM SPA sample screening capability without the need to procure costly specialised Cryo-TEMs. This could help to considerably lower the economic entry barrier for cryo-EM SPA and contribute to the "democratisation" of cryo-EM.

Matching journals

The top 3 journals account for 50% of the predicted probability mass.

50% of probability mass above

"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.