The Effect of Electrode-Tissue Distance on Unipolar and Bipolar Voltage Electrograms for large-Tip, Ring-, and Mini-Electrodes
Schlageter, V.; Luca, A.; Badertscher, P.; Krisai, P.; Kueffer, T.; Spreen, D.; Katic, J.; Osswald, S.; Schaer, B.; Sticherling, C.; Kühne, M.; Knecht, S.
Show abstract
BackgroundThe characteristics of intracardiac unipolar and bipolar voltage electrograms (EGM) acquired by electrophysiological catheters depend on the electrode design and configuration.The aim of the study was to assess the impact of electrode design and distance from the myocardial electric source on the unipolar and bipolar intracardiac electrograms recorded with a multi-electrode ablation catheter do deduce a cut-off for far-field versus near-field discrimination. MethodsWe retrospectively analyzed left atrial electroanatomical maps of 25 patients performed using an ablation catheter with a 4.5 mm tip-, mini- and 2 mm ring electrodes. The unipolar and bipolar EGMs were characterized based on peak-to-peak amplitude, signal duration (width), maximal slope, and relative power of the high frequency spectrum (HF_rel). Distances of the electrode from the tissue were calculated from the electroanatomic reconstruction. ResultsWe analyzed EGMs of 5183 catheter positions. The unipolar EGM of ring electrodes showed an increased amplitude (140%), slope (150%) and HF_rel (16% vs 11%) compared to the tip- and mini-electrodes. In contrast, for bipolar EGM, the tip-ring pair showed the largest amplitude, width, and slope. The median amplitude, slope, and HF_rel for the ring electrodes followed a power-law decay function with distance. A cut-off of 4 mm was determined for far-field versus near-field discrimination. ConclusionsWe showed a higher unipolar amplitude for small ring-electrodes compared to larger tip electrodes. Furthermore, a rapid decay of the amplitude, slope, and HF_rel with distance could be observed. The decay functions are suggestive for a near-field cut-off distance below 4mm from the tissue.
Matching journals
The top 4 journals account for 50% of the predicted probability mass.
Similar papers in this journal
Similar papers in this journal
- Peri-Procedural Intravascular Hemolysis during Atrial Fibrillation Ablation: A Comparison of Pulsed-Field with Radiofrequency Ablation 95%
- Enhancing Ventricular Tachycardia Ablation Outcomes: The Impact Of Functional Mapping In Chagas Cardiomyopathy 95%
- Multicenter analysis of atrioesophageal fistula rates before and after adoption of active esophageal cooling during atrial fibrillation ablation 94%
Similar papers in this journal
- Digitizing ECG image: new fully automated method and open-source software code 94%
- BRAVEHEART: Open-source software for automated electrocardiographic and vectorcardiographic analysis 93%
- Automated IntraVascular UltraSound Image Processing and Quantification of Coronary Artery Anomalies: The AIVUS-CAA software 91%
Similar papers in this journal
Similar papers in this journal
- Atrial functional substrates for the prediction of atrial fibrillation recurrence after pulmonary vein isolation 95%
- Investigating Electrocardiographic Abnormalities in Patients with Coronary Microvascular Dysfunction 91%
- Prognostic value of compact myocardial thinning in patients with left ventricular non-compaction 90%
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.