Multicentre diagnostic evaluation of OnSite COVID-19 Rapid Test (CTK Biotech) among symptomatic individuals in Brazil and The United Kingdom
Thompson, C. R.; Torres, P. M.; Kontogianni, K.; Bengey, D.; Byrne, R. L.; Noguera, S. V.; Luna-Muschi, A. J.; Andrade, P. S.; dos Santos Barboza, A.; Nishikawara, M.; Body, R.; de Vos, M.; Escadafal, C.; Adams, E. R.; Costa, S. F.; Cubas Atienzar, A. I.
Show abstract
The COVID-19 pandemic has given rise to numerous commercially available antigen rapid diagnostic tests (Ag-RDTs). To generate and share accurate and independent data with the global community, multi-site prospective diagnostic evaluations of Ag-RDTs are required. This report describes the clinical evaluation of OnSite COVID-19 Rapid Test (CTK Biotech, California, USA) in Brazil and The United Kingdom. A total of 496 paired nasopharyngeal (NP) swabs were collected from symptomatic healthcare workers at Hospital das Clinicas in Sao Paulo, and 211 NP swabs were collected from symptomatic participants at a COVID-19 drive-through testing site in Liverpool, England. These swabs were analysed by Ag-RDT and results were compared to RT-qPCR. The clinical sensitivity of the OnSite COVID-19 Rapid test in Brazil was 90.3% [95% Cl 75.1 - 96.7%] and in the United Kingdom was 75.3% [95% Cl 64.6 - 83.6%]. The clinical specificity in Brazil was 99.4% [95% Cl 98.1 - 99.8%] and in the United Kingdom was 95.5% [95% Cl 90.6 - 97.9%]. Analytical evaluation of the Ag-RDT was assessed using direct culture supernatant of SARS-CoV-2 strains from Wild-Type (WT), Alpha, Delta, Gamma and Omicron lineages. Analytical limit of detection was 1.0x103 pfu/mL, 1.0x103 pfu/mL, 1.0x102 pfu/mL, 5.0x103 pfu/mL and 1.0x103 pfu/mL, giving a viral copy equivalent of approximately 2.1x105 copies/mL, 2.1x104 copies/mL, 1.6x104 copies/mL, 3.5x106 copies/mL and 8.7 x 104 for the Ag-RDT, when tested on the WT, Alpha, Delta, Gamma and Omicron lineages, respectively. This study provides comparative performance of an Ag-RDT across two different settings, geographical areas, and population. Overall, the OnSite Ag-RDT demonstrated a lower clinical sensitivity than claimed by the manufacturer... Sensitivity and specificity from the Brazil study fulfilled the performance criteria determined by the World Health Organisation but the performance obtained from the UK study failed to. Further evaluation of the use of Ag-RDTs should include harmonised protocols between laboratories to facilitate comparison between settings.
Matching journals
The top 8 journals account for 50% of the predicted probability mass.
Similar papers in this journal
- Performance Evaluation Of A SARS-COV-2 Rapid Antigentest: Test Performance In The Community In The Netherlands 97%
- A prospective diagnostic evaluation of accuracy of self-taken and healthcare worker-taken swabs for rapid COVID-19 testing 97%
- Diagnostic accuracy of two commercial SARS-CoV-2 Antigen-detecting rapid tests at the point of care in community-based testing centers 97%
Similar papers in this journal
- Evaluation of commercial qPCR kits for detection of SARS-CoV-2 in pooled samples 95%
- Implementation of a Rapid RT-LAMP Saliva-based SARS-CoV-2 Testing Program in the Workplace 94%
- Evaluation and modelling of the performance of an automated SARS-CoV-2 antigen assay according to sample type, target population and epidemic trends 94%
Similar papers in this journal
- Performance and Implementation Evaluation of the Abbott BinaxNOW Rapid Antigen Test in a High-throughput Drive-through Community Testing Site in Massachusetts 96%
- Diagnostic accuracy of SARS-CoV-2 rapid antigen detection testing in symptomatic and asymptomatic children in the clinical setting 96%
- SARS-CoV-2 testing in the community: Testing positive samples with the TaqMan SARS-CoV-2 Mutation Panel to find variants in real-time 95%
Similar papers in this journal
- Field evaluation of a rapid antigen test (Panbio™ COVID-19 Ag Rapid Test Device) for the diagnosis of COVID-19 in primary healthcare centers 97%
- Self-testing for the detection of SARS-CoV-2 infection with rapid antigen tests 96%
- SARS-CoV-2 N gene dropout and N gene Ct value shift as indicator for the presence of B.1.1.7 lineage in a widely used commercial multiplex PCR assay 96%
Similar papers in this journal
- Age significantly influences the sensitivity of SARS-CoV-2 rapid antibody assays 96%
- Performance of SARS-CoV-2 rapid antigen test compared with real-time RT-PCR in asymptomatic individuals 95%
- Self-collected oral, nasal and saliva samples yield sensitivity comparable to professional-collected oro-nasopharyngeal swabs in SARS-CoV-2 diagnosis 95%
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.