Validity of freehand three-dimensional ultrasound system in measurement of three-dimensional surface shape of shoulder muscles

Umehara, J.; Fukuda, N.; Konda, S.; Hirashima, M.

2022-04-10 bioengineering
10.1101/2021.11.18.468912 bioRxiv
Show abstract

Freehand three-dimensional ultrasound (3DUS) system is a promising technique for accurately assessing muscle morphology. However, its accuracy has been validated mainly in terms of volume by examining lower limb muscles. This study aimed to validate 3DUS in the measurements of 3D surface shape and volume by comparing them with MRI measurements while ensuring the reproducibility of participant posture by focusing on the shoulder muscles. The supraspinatus, infraspinatus, and posterior deltoid muscles of 10 healthy males were scanned using 3DUS and MRI while secured by an immobilization support customized for each participant. A 3D surface model of each muscle was created from the 3DUS and MRI methods, and the agreement between them was assessed. For the muscle volume, the mean difference between the two models was within -0.51 cm3. For the 3D surface shape, the distances between the closest points of the two models and Dice similarity coefficient were calculated. The results showed that the median surface distance was less than 1.21 mm and Dice similarity coefficient was larger than 0.85. These results suggest that, given the above error is permitted, 3DUS can be used as an alternative to MRI in measuring volume and surface shape, even for the shoulder muscles.

Matching journals

The top 7 journals account for 50% of the predicted probability mass.

50% of probability mass above

"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.