Clinical evaluation of a RT-LAMP SARS-CoV-2 test for the Point-Of-Care, rapid, low-cost, integrating sample solid phase extraction and on which reagents are lyophilized.

Coz, E.; Garneret, P.; Martin, E.; Freitas do Nascimento, D.; Vilquin, A.; Hoinard, D.; Feher, M.; Grassin, Q.; Vanhomwegen, J.; Manuguerra, J.-C.; Mukherjee, S.; Olivo-Marin, J.-C.; Brient-Litzler, E.; Merzoug, M.; collin, e.; Tabeling, P.; rossi, b.

2021-10-04 infectious diseases
10.1101/2021.10.03.21264480 medRxiv
Show abstract

ObjectivesDetermine the sensitivity and specificity of a Point-Of-Care test ( COVIDISC) for SARS-COV2. The novelty of the test is to integrate, on the same (low-cost) compact plastic/paper device, solid phase RNA extraction and RT-LAMP amplification, all reagents being freeze-dried on it. MethodRetrospective study with a cohort of 99 patients characterized by real-time RT-PCR. The 37 positive naso-pharyngeal samples cover a broad range of viral loads (from 5 gc /{micro}L to 2 106 gc/ {micro}L of sample). ResultsThe COVIDISC found 36 positives (out of 37 by IP4 RT-PCR protocols) and 63 negatives (out of 62 by RT-PCR). ConclusionThe sensitivity of the COVIDISC, found in this 99-patient retrospective study, is 97% and the specificity 100%.