Combined Ventilation Of Two Subjects With A Single Mechanical Ventilator Using A New Medical Device: An In Vitro Study
Lugones, I.; Ramos, M.; Biancolini, M. F.; Orofino Giambastiani, R. E.
Show abstract
INTRODUCTIONThe SARS-CoV2 pandemic has created a sudden lack of ventilators. DuplicAR(R) is a novel device that allows simultaneous and independent ventilation of two subjects with a single ventilator. The aims of this study are: a) to determine the efficacy of DuplicAR(R) to independently regulate the peak and positive-end expiratory pressures in each subject, both under pressure-controlled ventilation and volume-controlled ventilation, and b) to determine the ventilation mode in which DuplicAR(R) presents the best performance and safety. MATERIALS AND METHODSTwo test lungs are connected to a single ventilator using DuplicAR(R). Three experimental stages are established: 1) two identical subjects, 2) two subjects with the same weight but different lung compliance, and 3) two subjects with different weight and lung compliance. In each stage, the test lungs are ventilated in two ventilation modes. The positive-end expiratory pressure requirements are increased successively in one of the subjects. The goal is to achieve a tidal volume of 7 ml/kg for each subject in all different stages through manipulation of the ventilator and the DuplicAR(R) controllers. RESULTSDuplicAR(R) allows adequate ventilation of two subjects with different weight and/or lung compliance and/or PEEP requirements. This is achieved by adjusting the total tidal volume for both subjects (in volume-controlled ventilation) or the highest peak pressure needed (in pressure-controlled ventilation) along with the basal positive-end expiratory pressure on the ventilator, and simultaneously manipulating the DuplicAR(R) controllers to decrease the tidal volume or the peak pressure in the subject that needs less and/or to increase the positive-end expiratory pressure in the subject that needs more. While ventilatory goals can be achieved in any of the ventilation modes, DuplicAR(R) performs better in pressure-controlled ventilation, as changes experienced in the variables of one subject do not modify the other one. CONCLUSIONSDuplicAR(R) is an effective tool to manage the peak inspiratory pressure and the positive-end expiratory pressure independently in two subjects connected to a single ventilator. The driving pressure can be adjusted to meet the requirements of subjects with different weight and lung compliance. Pressure-controlled ventilation has advantages over volume-controlled ventilation and is therefore the recommended ventilation mode.
Matching journals
The top 1 journal accounts for 50% of the predicted probability mass.
Similar papers in this journal
- Sharing ventilators in the Covid-19 pandemics. A bench study 98%
- PVP1–The People’s Ventilator Project: A fully open, low-cost, pressure-controlled ventilator 95%
- Computational simulation to assess patient safety of uncompensated COVID-19 two-patient ventilator sharing using the Pulse Physiology Engine 95%
Similar papers in this journal
- Comparing physiological impacts of positive pressure ventilation versus self-breathing via a versatile cardiopulmonary model incorporating a novel alveoli opening mechanism 95%
- Quantifiable identification of flow-limited ventilator dyssynchrony with the deformed lung ventilator model 92%
- Improving irregular temporal modeling by integrating synthetic data to the electronic medical record using conditional GANs: a case study of fluid overload prediction in the intensive care unit 91%
Similar papers in this journal
- Performance of EasyBreath ® Decathlon Snorkeling mask for Delivering Continuous Positive Airway Pressure 97%
- A Cross-sectional Feasibility Study to Evaluate the Usability and Efficacy of “Swaasa”: An AI-Platform for Rapid Respiratory Health Assessment 94%
- Newly developed silicone airway stent (GINA stent): Mechanical characteristics and performance evaluation in pigs 92%
Similar papers in this journal
- De novo Powered Air-Purifying Respirator Design and Fabrication for Pandemic Response 90%
- A Increased trunk flexion in standing up is related to muscle weakness rather than pain avoidance in individuals with unilateral knee pain; a simulation study 87%
- Hot, humid air decontamination of aircraft confirmed that high temperature and high humidity are critical for inactivation of infectious, enveloped ribonucleic acid (RNA) virus 86%
Similar papers in this journal
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.