A Rapidly Deployable Negative Pressure Enclosure for Aerosol-Generating Medical Procedures
Chahal, A. M.; Van Dewark, K.; Gooch, R.; Fukushima, E.; Husdon, Z. M.
Show abstract
BackgroundThe coronavirus disease 2019 (COVID-19) pandemic presents significant safety challenges to healthcare professionals. In some jurisdictions, over 10% of confirmed cases of COVID-19 have been found among healthcare workers. Aerosol-generating medical procedures (AGMPs) may increase the risk of nosocomial transmission, exacerbated by present global shortages of personal protective equipment (PPE). Improved methods for mitigating risk during AGMPs are therefore urgently needed. MethodsThe Aerosol Containment Enclosure (ACE) was constructed from acrylic with silicone gaskets for arm port seals and completed with a thin plastic sheet. Hospital wall suction generated negative pressure within the ACE. To evaluate protective capability, differential pressures were recorded under static conditions and during simulated AGMPs. Smoke flow patterns, fluorescence aerosolization, and sodium saccharin aerosolization tests were also conducted. ResultsNegative pressures of up to -47.7 mmH2O were obtained using the enclosure with two wall suction units (combined outflow of 70 L min-1), with inflow of O2 of 15 L min-1. Negative pressures between -10 and - 35 mmH2O were maintained during simulated AGMPs, including oxygen delivery by mask, airway suctioning, bag-mask manual ventilation and endotracheal intubation of a potential COVID-19 patient. The ACE effectively contained smoke, fluorescein aerosol, and sodium saccharin aerosol within the enclosure during use. ConclusionsThe ACE is capable of maintaining negative pressure during simulated AGMPs. In all cases, containment was improved relative to an identical enclosure with non-occluded ports at ambient pressure. During the current COVID-19 pandemic, the use of such a device may assist in reducing nosocomial infections among healthcare providers.
Matching journals
The top 3 journals account for 50% of the predicted probability mass.
Similar papers in this journal
- A simple, effective enclosure with disposable coverings for inexpensive containment of aerosolized COVID viruses during tracheal intubation and extubation 96%
- Exploring options for reprocessing of N95 Filtering Facepiece Respirators (N95-FFRs) amidst COVID-19 pandemic: a systematic review 96%
- Decontamination of Geobacillus Stearothermophilus Using the Arca Aerosolized Hydrogen Peroxide Decontamination System 95%
Similar papers in this journal
- Analysis of national and international guidelines on respiratory protection equipment for COVID-19 in healthcare settings. 95%
- Elastomeric Harnesses Decrease Aerosol Particles Leak and Improve the Fit of Surgical Masks 95%
- Estimating the Per-Application Effectiveness of Chlorhexidine Gluconate and Mupirocin in Methicillin-resistant Staphylococcus aureus Decolonization in Intensive Care Units 90%
Similar papers in this journal
- Improving communications in PPE: A solution for ‘landline’ telephone communication 94%
- Occupational Exposures and Programmatic Response to COVID-19 Pandemic: An Emergency Medical Services Experience 90%
- Diversity of CPR manikins for basic life support education: Use of manikin sex, race, and body shape – A scoping review 87%
Similar papers in this journal
- Effectiveness of portable air filtration on reducing indoor aerosol counts: preclinical observational trials 93%
- Exposition to airborne SARS-CoV-2 in household and hospitals settings as studied by a vacuum cleaner as a high-powered air sampler 93%
- Use, reuse or discard: quantitatively defined variance in N95 respirator integrity following vaporized hydrogen peroxide decontamination during the COVID-19 pandemic 93%
Similar papers in this journal
- Evaluation of at-home methods for N95 filtering facepiece respirator decontamination 93%
- Computational Fluid Dynamics Modeling of Cough Transport in an Aircraft Cabin 92%
- Scalable, effective, and rapid decontamination of SARS-CoV-2 contaminated N95 respirators using germicidal ultra-violet C (UVC) irradiation device 92%
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.