Biomechanical Compatibility With Physiological Limb-Flexion Deformation Reduces Inflammation and Neointimal Hyperplasia After Peripheral Stenting
Kamenskiy, A.; de Oliveira, B. B.; Desyatova, A.; Heinis, F.; Rosenbohm, J.; Deegan, P.; Renavikar, P.; MacTaggart, J.
Show abstract
Introduction: The femoropopliteal artery undergoes substantial axial compression, bending, torsion, and pinching during limb flexion. When an implanted stent does not accommodate these deformations, repetitive biomechanical mismatch may alter the vascular response to implantation and contribute to restenosis. We tested whether designing a peripheral stent to better accommodate physiologic limb-flexion deformation reduces the subsequent inflammatory and neointimal response. Methods: A nitinol stent was developed using multi-objective computational optimization to maintain arterial apposition while reducing wall stress and luminal pinching during physiologic deformation. Sixteen Yucatan minipigs were studied in independent 3- and 6-month cohorts (n=8 each). Each animal received the optimized stent in one limb and a commercial stent with representative biomechanical behavior contralaterally. CTA quantified in vivo limb-flexion deformation, while CT and histology assessed stenosis, neointimal accumulation, inflammatory-cell burden, SMA-positive tissue, and extracellular-matrix composition. Results: The optimized stent accommodated greater axial shortening than the commercial comparator at both 3 months (11.2{+/-}2.7% vs. 6.6{+/-}2.1%, p=0.003) and 6 months (13.7{+/-}4.4% vs. 6.8{+/-}2.3%, p=0.004). Intimal hyperplasia volume was lower with the optimized stent at 3 months (227.5{+/-}85.2 vs. 350.0{+/-}130.5 mm3, p=0.002) and 6 months (183.6{+/-}46.2 vs. 261.2{+/-}60.6 mm3, p=0.001), with significantly lower maximum, mean, and median stenosis at both timepoints. At 3 months, macrophage and total inflammatory-cell counts were 51% and 45% lower, respectively, with the optimized stent. Histologic tissue area was 28% lower at 3 months and 12% lower at 6 months. SMA-positive tissue fraction was higher with the optimized stent at both timepoints, whereas elastin, glycosaminoglycan, and collagen fractions were similar between devices. Discussion: Improved accommodation of limb-flexion deformation was associated with a smaller inflammatory and neointimal response to stent implantation through 6 months, while overall neointimal matrix composition remained largely unchanged. These findings provide in vivo evidence that biomechanical compatibility with the dynamic lower-limb environment can influence vascular healing and should be considered a mechanistic design criterion for peripheral artery stents.
Matching journals
The top 7 journals account for 50% of the predicted probability mass.
Similar papers in this journal
- Structural and Mechanical Analysis of Treated and Untreated Aortic Coarctation in a Growing Porcine Model 94%
- An investigation into the critical role of fibre orientation in the ultimate tensile strength and stiffness of human carotid plaque caps 92%
- Advanced Manufacturing of Coil-Reinforced Multilayer Vascular Grafts to Optimize Biomechanical Performance 92%
Similar papers in this journal
- Biomechanical Characterization Of Porcine Lower Limb Arteries For Preclinical Evaluation Of Peripheral Vascular Devices 97%
- Endoluminal catheter pulsed field ablation for the treatment of atherosclerotic vascular disease 93%
- Effect of Sinotubular Junction Size on TAVR Leaflet Thrombosis: A Fluid-structure Interaction Analysis 92%
Similar papers in this journal
- Improved Hemodynamic Performance and Reduced Paravalvular Regurgitation with the SAPIEN 3 Ultra RESILIA Valve: A Propensity-Matched Single-Center TAVR Study 94%
- Autologous cardiac micrografts as support therapy to coronary artery bypass surgery 91%
- Predicting long-term prognosis after percutaneous coronary intervention in patients with acute coronary syndromes: a prospective nested case-control analysis for county-level health services 91%
Similar papers in this journal
- Experiments and Simulations to Assess Exercise-Induced Pressure Drop Across Aortic Coarctations 92%
- Stretching the Limits: From Planar-Biaxial Stress-Stretch to Arterial Pressure-Diameter 90%
- Remodeling-mediated changes in left ventricular mechanics under settings of chronic pressure overload and exercise 89%
Similar papers in this journal
- In vivo recellularization of xenogeneic vascular grafts decellularized with high hydrostatic pressure method in a porcine carotid arterial interpose model 94%
- Effects of low versus high inspired oxygen fraction on myocardial injury after transcatheter aortic valve implantation: A randomized clinical trial 92%
- Full-field optical coherence tomography for the diagnosis of giant cell arteritis 91%
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.