MR-ARFI and Beyond: Characterizing the Temporal Dynamics of Acoustic Radiation Force in Human Brain
Mohammadjavadi, M.; Pauly, K. B.; Glover, G. H.
Show abstract
Purpose: Magnetic Resonance Acoustic Radiation Force Imaging (MR-ARFI) encodes focused ultrasound (FUS)-induced tissue displacements to MR phase shifts. Because these phase shifts are small, optimizing displacement sensitivity and minimizing motion artifacts requires characterization of the temporal dynamics of acoustic radiation force (ARF). Beyond sequence design, such characterization enables MR-ARFI to serve as a probe of brain viscoelastic behavior and can inform the design of therapeutic ultrasound pulsing schemes by quantifying how rapidly tissue deforms and relaxes. Methods: We developed a dynamic model assuming a gamma-variate impulse response for ARF(t), with time constant ({tau}) to be estimated. A single-shot spin echo ARFI acquisition with a bipolar motion encoding gradient (MEG) was implemented, incorporating retrospective corrections for head motion and physiological noise. A 500 kHz FUS pulse was applied in on/off blocks while systematically varying the timing between the MEG and sonication. The resulting phase time series were fit to the model to estimate time constant in five human participants. To assess sensitivity to tissue stiffness, {tau} was additionally measured in two customized gelatin phantoms of differing stiffness, characterized independently by MR elastography (MRE). Results: The average measured time constant in human participants was {tau} 3.7 {+/-} 0.27 ms. Comparing the two phantoms, the stiffer phantom exhibited a shorter {tau} (3.0-3.5 ms) than the softer phantom (4.0-4.5 ms), which tracked the MRE-derived Kelvin-Voigt viscoelastic time constants. In addition, the peak ARFI displacement was lower in the stiffer phantom (0.17 vs 0.74 m for the less stiff phantom), and a broader displacement focal spot size (FWHM 22 mm vs 15 mm). Conclusion: This time series approach with motion and physiological noise corrections validates an MR-based technique capable of characterizing the brain's temporal response to sonication-induced mechanical forces, providing a foundation for optimizing the design of ultrasound pulsing schemes for ARFI, as well as for future studies of brain biomechanics.
Matching journals
The top 3 journals account for 50% of the predicted probability mass.
Similar papers in this journal
- Mesoscopic whole-brain T2*-weighted and associated quantitative MRI in healthy humans at 10.5 T 95%
- TRACC-PHYSIO: Time-domain Resolution-Aligned Cross-Correlation to estimate PHYSIOlogical coupling and time delays in dynamic MRI 95%
- Robust 3D Bloch-Siegert based B1+ mapping using Multi-Echo General Linear Modelling 95%
Similar papers in this journal
- Multi-modal Monte Carlo MRI simulator of tissue microstructure 95%
- Robust functional ultrasound imaging in the awake and behaving brain: a systematic framework for motion artifact removal 94%
- Visual stimulus-evoked blood velocity responses in individual human posterior cerebral arteries measured with dynamic phase-contrast functional MR angiography 94%
Similar papers in this journal
Similar papers in this journal
- A Phantom System Designed to Assess the Effects of Membrane Lipids on Water Proton Relaxation 94%
- Deep-Learning-Based Accelerated and Noise-Suppressed Estimation (DANSE) of quantitative Gradient Recalled Echo (qGRE) MRI metrics associated with Human Brain Neuronal Structure and Hemodynamic Properties 94%
- Hierarchical Bayesian Modelling Improves Microstructural Parameter Mapping in Diffusion and Exchange MRI Data 93%
Similar papers in this journal
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.