A harmonized phantom MRI quality control framework identifies sources of longitudinal and multi-site variability
Yague Jimenez, B.; Kalantari, A.; Bolster, D.; Botto, E.; Carnevale, L.; Doblas, S.; Gammaraccio, F.; Gottschalk, M.; Hekmatyar, K.; Klohs, J.; Kurniawan, N.; Lopez-Gil, X.; Maus, B.; Micotti, E.; Moro, F.; Mueller, S.; Perez Nieto, L.; Po, C.; Swanberg, K.; Trotier, A.; Van Audekerke, J.; Van der Toorn, A.; Van Spilbeeck, I.; Wenk, P.; Basse-Luesebrink, T.; Bertoglio, D.; Boehm-Sturm, P.; Budinger, E.; das Neves Duarte, J. M.; Dijkhuizen, R.; Faber, C.; Forloni, G.; Garteiser, P.; Holmes, W.; Longo, D.; Meier, M.; Munoz-Moreno, E.; Ribot, E.; Vasco Simoes, R.; Ielacqua, G.; Aswendt, M.
Show abstract
The shift towards open science in preclinical research requires high-quality, comparable data adhering to FAIR principles, yet rigorous quality assurance (QA) and quality control (QC) frameworks remain less established in preclinical magnetic resonance imaging (MRI) than in clinical imaging. To address this limitation, a multicenter study was conducted across 21 international laboratories using standardized commercial liquid phantoms for mouse and rat MRI setups and a harmonized acquisition protocol. Data processing was centralized using AIDAqc, an automated pipeline extracting quantitative metrics including signal-to-noise ratio (SNR), temporal SNR (tSNR), ghost-to-signal ratio (GSR), and motion-equivalent temporal instability, combined with five complementary outlier-detection algorithms. The evaluated datasets encompassed magnetic field strengths from 3.0 to 16.4 T and heterogeneous coil and acquisition configurations. In mouse phantom data, SNR and tSNR were significantly lower at 3 T than at 7 and 9.4 T, whereas artifact-related metrics did not differ significantly between field-strength groups. In rat phantom data, differences between field strengths were less pronounced. At 7 T, substantial dataset-specific differences were observed in anatomical and functional quality metrics, also among datasets using the similar coil configuration. Multicenter reference values for frequently represented 7 T configurations were established, including SNR/tSNR of 48 {+/-} 3/48 {+/-} 1 dB for mouse surface-coil datasets and 49 {+/-} 3/44 {+/-} 4 dB for rat array-coil datasets. Longitudinal anatomical SNR showed generally low variability, whereas ghosting was more variable across time. Automated outlier detection additionally identified transient acquisition instabilities that were not always apparent by visual inspection. Together, these findings show that field strength and nominal coil configuration alone do not adequately characterize MRI performance and support standardized, longitudinal phantom-based QC combined with automated analysis as a practical approach for improving the reliability, transparency, and interoperability of multicenter preclinical MRI data.
Matching journals
The top 1 journal accounts for 50% of the predicted probability mass.
Similar papers in this journal
- Mesoscopic whole-brain T2*-weighted and associated quantitative MRI in healthy humans at 10.5 T 96%
- A Numerical Alternative to MR Thermometry for Safety Validation of Multi-Channel RF Transmit Coils 96%
- Multi-center benchmarking of cervical spinal cord RF coils for 7 T MRI: A traveling spines study 96%
Similar papers in this journal
- A Phantom System Designed to Assess the Effects of Membrane Lipids on Water Proton Relaxation 95%
- Hierarchical Bayesian Modelling Improves Microstructural Parameter Mapping in Diffusion and Exchange MRI Data 94%
- Non-negative least squares computation for in vivo myelin mapping using simulated multi-echo spin-echo T2 decay data 94%
Similar papers in this journal
- Multi-modal Monte Carlo MRI simulator of tissue microstructure 94%
- Test-retest repeatability of intravoxel incoherent motion (IVIM) measurements in the cervical cord 94%
- Automated quality control of T1-weighted brain MRI scans for clinical research: methods comparison and design of a quality prediction classifier 92%
Similar papers in this journal
- The dot-compartment revealed? Diffusion MRI with ultra-strong gradients and spherical tensor encoding in the living human brain 94%
- Cerebrospinal Fluid Flow within Ventricles and Subarachnoid Space Evaluated by Velocity Selective Spin Labeling MRI 94%
- Increased Sensitivity and Signal-to-Noise Ratio in Diffusion-Weighted MRI using Multi-Echo Acquisitions 94%
Similar papers in this journal
- Increased brain coverage and efficiency when measuring current-induced magnetic fields by use of simultaneous multi-slice echo-planar MRI 96%
- Acceleration of chemical shift encoding-based water fat MRI for liver proton density fat fraction and T2* mapping using compressed sensing 94%
- Analysis of physiological noise in quantitative cardiac magnetic resonance 94%
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.