Targeted and bilateral blood flow monitoring in middle cerebral artery using diffuse correlation spectroscopy

Sharma, K.; Das, S.; Gonsalves, K.; Sarkar, S.; Srinivasan, U. S.; Varma, H. M.

2026-09-25 bioengineering
10.64898/2026.09.21.750817 bioRxiv
Show abstract

Objective: To develop and validate a dual-probe Diffuse Correlation Spectroscopy (DCS) system for non-invasive and simultaneous, monitoring of cerebral blood flow (CBF) in the bilateral Middle Cerebral Artery (MCA) territories, and expanding the utility of conventional DCS limited to cortical-volume-based CBF measurements to vessel-specific cerebral perfusion monitoring. Methods: A dual-probe DCS system was designed for non-invasive monitoring of MCA-specific perfusion. Probe placement and protocol optimization study has been performed using anatomical landmarks, motor and speech activation tasks in healthy volunteers. System stability and repeatability were further evaluated in a pilot cohort of 30 healthy (age, 25{+/-}7 years) participants using optimized probe position and protocol. A bilateral MCA ischemic Lacunar Infract stroke case report also validated the feasibility of the system in clinical settings. Results: Measurements demonstrated superior sensitivity towards MCA-territory perfusion at targeted probe locations compared to off-MCA positions. In pilot cohort, significant increase of 30.34 {+/-} 21.56% and 36.48 {+/-} 21.22% in rCBF corresponding to hand squeeze and speech task respectively showed reproducible physiological responsiveness of the system (p<0.001). Measurement done on a patient with bilateral MCA ischemic Lacunar Infract stroke showed a significant change of 30% during speech for both the MCAs but no significant change is observed for hand squeeze tasks. Conclusion: The custom built dual-probe DCS system enables non-invasive, operator-independent, targeted and continuous monitoring of rCBF within bilateral MCA territories. Significance: This approach enables the potential use of DCS system for bilateral and vessel-specific monitoring of cerebral perfusion in the MCA territories.

Matching journals

The top 10 journals account for 50% of the predicted probability mass.

50% of probability mass above

"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.