Deep Learning Reconstruction for Ultra-High-Resolution Photon-Counting CT of the Lung: Image Quality and Texture Characterization
Mei, K.; Roshkovan, L.; Liu, L. P.; Halliburton, S. S.; Sharma, S.; Ross, S.; Stroud, T.; Akino, N.; Yu, Z.; Thompson, R.; Dhanaliwala, A. H.; Litt, H. I.; Noel, P. B.
Show abstract
Objectives: To characterize image noise, lesion contrast, spatial resolution, and attenuation distribution for deep learning reconstruction (DLR) compared with hybrid iterative reconstruction (IR) in ultra-high-resolution (UHR) photon-counting CT (PCCT) of the lung, using a patient-derived 3D-printed phantom, and to apply the same analytical framework to patient examinations acquired on the same system. Materials and Methods: A patient CT scan-derived lung phantom containing three digitally inserted lesions of differing morphology was fabricated using PixelPrint 3D printing and imaged on a CZT-based PCCT prototype at four dose levels (CTDIvol 6, 4, 2, and 1 mGy). Images were reconstructed with hybrid iterative reconstruction at normal resolution (512 by 512) and UHR (1024 by 1024), and with DLR at UHR. Image noise, lesion contrast, and contrast-to-noise ratio (CNR) were measured within prespecified regions of interest replicated across reconstructions by coordinate transfer. Spatial resolution was assessed by the modulation transfer function (MTF) from a high-contrast edge at eight locations. Attenuation distribution and local texture were characterized using histograms and gray-level co-occurrence matrices (GLCMs) within a structured parenchymal region. Two patient examinations with part solid lung nodules acquired on the same prototype were analyzed as illustrative examples, with solid and ground-glass components segmented separately and their attenuation distributions compared. Results: In the phantom, DLR reduced image noise relative to UHR hybrid iterative reconstruction (7.8% at 6 mGy and 15.2% at 1 mGy). Lesion contrast was constant across reconstructions and dose levels, so CNR differences reflected differences in noise alone. MTF curves for DLR lay above those for hybrid iterative reconstruction across the evaluated spatial frequency range at both 6 and 1 mGy, averaged +0.4 lp/mm in MTF50. In a structured parenchymal region, the attenuation histogram mode became progressively narrower from hybrid iterative reconstruction at 1 mGy to DLR at 6 mGy (averaged -23%), while mode position and mean regional attenuation were preserved, consistent with reduced partial-volume mixing rather than altered attenuation. In the patient example, the attenuation distributions of solid and ground-glass components were more widely separated with UHR DLR (2.0) than with either hybrid iterative reconstruction (1.6). Conclusions: In UHR photon-counting CT of the lung, DLR reduced image noise while preserving lesion contrast and regional attenuation, increased measured MTF at a high-contrast edge, and produced attenuation distributions consistent with reduced partial-volume mixing at material boundaries. The same effect was observed in the separation of solid and ground-glass components in two patient examples.
Matching journals
The top 4 journals account for 50% of the predicted probability mass.
Similar papers in this journal
- Patient-derived PixelPrint phantoms for evaluating clinical imaging performance of a deep learning CT reconstruction algorithm 97%
- Reproducible spectral CT thermometry with liver-mimicking phantoms for image-guided thermal ablation 95%
- Contrast-enhanced photon-counting micro-CT of tumor xenograft models 93%
Similar papers in this journal
- Inconsistency of AI in Intracranial Aneurysm Detection with Varying Dose and Image Reconstruction 93%
- Identifying relationships between imaging phenotypes and lung cancer-related mutation status: EGFR and KRAS 92%
- High-Dimensional Multinomial Multiclass Severity Scoring of COVID-19 Pneumonia Using CT Radiomics Features and Machine Learning Algorithms 91%
Similar papers in this journal
- First-generation clinical dual-source photon-counting CT: ultra-low dose quantitative spectral imaging 97%
- Impact of Non-Contrast Enhanced Imaging Input Sequences on the Generation of Virtual Contrast-Enhanced Breast MRI Scans using Neural Networks 92%
- From Community Acquired Pneumonia to COVID-19: A Deep Learning Based Method for Quantitative Analysis of COVID-19 on thick-section CT Scans 92%
Similar papers in this journal
- Fully Automated Explainable Abdominal CT Contrast Media Phase Classification Using Organ Segmentation and Machine Learning 93%
- PSMA-Hornet: fully-automated, multi-target segmentation of healthy organs in PSMA PET/CT images 93%
- Phase Recognition in Contrast-Enhanced CT Scans based on Deep Learning and Random Sampling 92%
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.