Combining Clinical LAFOV PET/CT with a Digital Twin Providing Motion-Free Ground Truth Reveals Quantitative Trade-offs in Respiratory Motion Correction

Lan, W.; Weigel, S.; Calderon, E.; Fougere, C. l.; Schmidt, F. P.

2026-08-12 radiology and imaging
10.64898/2026.08.11.26360175 medRxiv
Show abstract

Purpose: Respiratory motion remains a major source of quantitative bias in PET and becomes increasingly relevant for high-sensitivity long axial field-of-view (LAFOV) PET/CT. Although numerous respiratory motion correction (MoCo) methods have been proposed, their quantitative accuracy cannot be established clinically because a patient-specific motion-free reference is fundamentally unavailable in vivo. This study combined clinical PET imaging with a digital twin, a realistic representation of both the PET/CT system and the patient, to objectively validate respiratory MoCo against a corresponding motion-free reference. Methods: Twenty patients (10 [18F]FDG with predominantly pulmonary lesions and 10 [18F]SiFAlin-TATE with predominantly hepatic lesions; total 135 lesions) were analyzed. The digital twin combined a validated LAFOV PET/CT simulation model with an anatomically realistic phantom containing 14 lung and liver lesions, two patient-derived respiratory patterns, and respiratory motion amplitudes of 2 and 3 cm, generating patient-like datasets with corresponding motion-free references. Data-driven and image-based MoCo were evaluated using lesion morphology, SUVmean, SUVmax, and metabolic tumor volume (MTV). Results: In patients, data-driven MoCo produced larger SUVmean increases than image-based MoCo for liver (48.1{+/-}18.9% vs. 17.0 {+/-} 12.0%; p<0.01), lower-lung (32.5{+/-}21.2% vs. 16.3{+/-}15.6%, p=0.06), and upper-lung lesions (28.4{+/-}32.0% vs. 10.4 {+/-} 17.2%; p<0.01), with similar findings for SUVmax and larger MTV reductions. Simulation revealed marked motion-induced SUVmean underestimation before correction, particularly in liver (-31.2{+/-}6.8%) and lower lung (-15.5{+/-}13.9%). Relative to the motion-free reference, data-driven MoCo most accurately recovered hepatic uptake (4.3{+/-}11.7% vs. -10.0 {+/-} 9.2%; p=0.01) but overestimated pulmonary uptake (lower lung: 19.8{+/-}16.3% vs. -1.6 {+/-} 10.2%; p=0.02). SUVmax showed the same regional behavior, whereas image-based MoCo yielded MTV estimates closer to the reference. Quantitative recovery was largely independent of respiratory pattern, while larger motion amplitudes mainly affected image-based MoCo. Conclusion: Combining clinical PET with a realistic digital twin and corresponding motion-free ground truth enabled objective validation of respiratory MoCo beyond conventional clinical evaluation. Larger correction-induced quantitative changes should not be equated with greater quantitative accuracy. Instead, MoCo performance was region- and metric-dependent, highlighting the value of ground-truth-based validation for developing and benchmarking respiratory motion correction and quantitative PET on LAFOV PET/CT systems.

Matching journals

The top 5 journals account for 50% of the predicted probability mass.

50% of probability mass above

"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.