Dual-Filament 3D Printing of Patient-Specific CT Phantoms with Embedded Implants and Tunable Metal-Artifact Intensity

Pasyar, P.; Mei, K.; Im, J. Y.; Roshkovan, L.; Geagan, M.; Noël, P. B.

2026-07-20 radiology and imaging
10.64898/2026.07.17.26358319 medRxiv
Show abstract

ABSTRACT Background: Metallic implants such as orthopedic screws, prostheses, and dental hardware produce beam-hardening, photon-starvation, and streak artifacts that degrade computed tomography (CT) image quality, and the metal artifact reduction (MAR) methods developed to mitigate them require objective, reproducible benchmarking. Purpose: Objective evaluation of MAR algorithms in CT is hindered by the absence of phantoms that simultaneously provide anatomically realistic backgrounds, embedded implants of known geometry, and controllable, ground-truth--referenced artifact intensity. We present a dual-filament, voxel-level three-dimensional (3D) printing method that fulfills these requirements and demonstrate its capabilities on a clinically representative cervical spine case with embedded orthopedic spinal screws. Methods: The proposed method extends the PixelPrint framework, a fused-deposition-modeling (FDM) workflow that converts clinical Digital Imaging and Communications in Medicine (DICOM) data directly into 3D-printer Geometric code (G-code) without intermediate segmentation or surface meshing, to interleaved, voxel-level deposition of two filaments: a calcium-doped polylactic acid (PLA) for soft tissue and bone, and a higher-attenuation metal-doped PLA for metallic implants. For demonstration, anonymized DICOM data of a healthy cervical spine were used to design and fabricate three matched phantoms, each with six embedded spinal screws at C4--C6: a 0% metal-infill ground-truth phantom, a 50% medium-metal-infill phantom, and an 85% high-metal-infill phantom. All phantoms were scanned on a clinical spectral CT system at 120 kVp and 1000 mAs, reconstructed at 0.67 mm slice thickness with virtual monoenergetic imaging (VMI) across 50--190 keV. Method performance was characterized by region of interest (ROI)-based Hounsfield Unit (HU) agreement with the source patient data and by the noise-independent Gumbel-distribution p-index metric. Results: The dual-filament method reproduced patient anatomy, soft-tissue contrast, and screw geometry with high fidelity. ROI HU values agreed with patient data within {+/-}25 HU for soft tissue and trabecular bone; cortical regions were underestimated owing to the current ceiling of the calcium-doped PLA used in this study. The tunable-artifact behavior was quantified as follows: the Gumbel location parameter scaled monotonically from 46.7 HU (no-metal background) to 57.1 HU (50% infill) to 90.5 HU (85% infill) for the VMI 70 keV with standard filter. High-keV VMI reconstructions substantially reduced streak and beam-hardening artifacts while preserving anatomic detail. Conclusions: The proposed dual-filament, voxel-level PixelPrint method enables the fabrication of patient-specific, multi-material CT phantoms with embedded metallic implants and controllable, ground-truth--referenced artifact intensity. Although demonstrated here in a single cervical-spine case, the workflow is anatomy- and implant-agnostic by construction and could in principle be adapted to other musculoskeletal sites (e.g., knee, hip, dental) and implant materials, providing a reproducible methodological foundation for benchmarking MAR algorithms, characterizing spectral CT performance, and validating emerging photon-counting detector systems. Keywords: 3D printing methodology; fused deposition modeling; voxel-level multi-material printing; spectral computed tomography; metal artifact reduction; phantom design; orthopedic implants; dual filament; PixelPrint.

Matching journals

The top 5 journals account for 50% of the predicted probability mass.

1
Physics in Medicine & Biology
18 papers in training set
Top 0.1%
18.8%
2
Medical Physics
14 papers in training set
Top 0.1%
12.9%
3
European Radiology
15 papers in training set
Top 0.1%
9.9%
4
Scientific Reports
3612 papers in training set
Top 10%
6.8%
5
PLOS ONE
5266 papers in training set
Top 28%
5.6%
50% of probability mass above
6
Imaging Neuroscience
282 papers in training set
Top 1%
5.2%
7
Biomedical Optics Express
95 papers in training set
Top 0.3%
4.1%
8
Bone
25 papers in training set
Top 0.2%
2.8%
9
Journal of Medical Imaging
11 papers in training set
Top 0.1%
2.2%
10
Annals of Biomedical Engineering
37 papers in training set
Top 0.5%
1.9%
11
Science Advances
1243 papers in training set
Top 18%
1.9%
12
Magnetic Resonance in Medicine
85 papers in training set
Top 0.4%
1.8%
13
GigaScience
212 papers in training set
Top 2%
1.8%
14
Biomedical Physics & Engineering Express
11 papers in training set
Top 0.2%
1.5%
15
International Journal of Radiation Oncology*Biology*Physics
25 papers in training set
Top 0.3%
1.5%
16
Nature Communications
5641 papers in training set
Top 49%
1.4%
17
Photoacoustics
12 papers in training set
Top 0.2%
1.1%
18
Archives of Clinical and Biomedical Research
28 papers in training set
Top 0.9%
1.0%
19
NeuroImage
903 papers in training set
Top 6%
0.9%
20
Computers in Biology and Medicine
128 papers in training set
Top 4%
0.9%
21
Frontiers in Oncology
103 papers in training set
Top 3%
0.9%
22
Radiotherapy and Oncology
19 papers in training set
Top 0.3%
0.9%
23
IEEE Journal of Biomedical and Health Informatics
37 papers in training set
Top 1%
0.9%
24
Journal of Magnetic Resonance Imaging
16 papers in training set
Top 0.4%
0.6%