Clinical-grade Cuffless Blood Pressure Monitoring via Deep-tissue Diffuse Speckle Pulsatile Flowmetry
Choo, T. W. J.; Yap, A. A. M.; Kawaja, A.; Chao, V. T. T.; Ng, C. J.; Olivo, M.; Bi, R.
Show abstract
Blood pressure (BP) is a vital sign which is measured to diagnose and manage hypertension. However, current methods to measure BP use inflatable cuffs which cause discomfort and limit the frequency at which measurements can be made, or intra-arterial catheters which are invasive and pose infection risks. Here, we propose and evaluate the use of Diffuse Speckle Pulsatile Flowmetry (DSPF) as a cuffless BP measurement method to address these limitations. DSPF is a laser speckle-based technique which simultaneously records blood flow rate and blood volume (i.e. photoplethysmography or PPG) signals from relatively deep vascular tissue. Using information from these signals, we studied DSPFs effectiveness in measuring systolic BP (SBP) and diastolic BP (DBP) through an outpatient study in which 133 patients were recruited, and in measuring beat-to-beat BP waveforms through an inpatient study in which two patients were recruited. In the outpatient study, the DSPF method was able to achieve mean absolute errors (MAEs) of 4.17 mmHg and 2.42 mmHg for SBP and DBP respectively compared to conventional cuff-based methods. It was also able to fulfil the requirements of the AAMI/ESH/ISO 81060-2:2018 standard for BP measurement devices and attain an "A" grade according to the British Hypertension Society grading scheme. For the inpatient study, it produced BP waveforms which had MAEs of 2.35 mmHg and 3.06 mmHg compared to arterial-line measurements for the two patients, respectively. Compared to PPG which has been studied more extensively as a cuffless BP measurement method, we found through ablation studies that DSPF was able to reach significantly lower MAEs and hence better accuracies. DSPF augments the performance of PPG-only methods by leveraging additional information from the blood flow rate signal, and we therefore find it to be a superior cuffless BP measurement method which can potentially be used in outpatient, inpatient, and remote settings.
Matching journals
The top 7 journals account for 50% of the predicted probability mass.
Similar papers in this journal
- Coarctation duration and severity predict risk of hypertension precursors in a preclinical model and hypertensive status among patients 92%
- Sympathetic Vasomotion Reflects Catheter-based Radiofrequency Renal Denervation 91%
- Blood-brain barrier integrity decreases with higher blood pressure, a 7T DCE-MRI study 91%
Similar papers in this journal
- Algorithm for Predicting Valvular Heart Disease from Heart Sounds in an Unselected Cohort 94%
- Deep learning-based end-to-end automated stenosis classification and localization on catheter coronary angiography 93%
- Fairness in Cardiac Magnetic Resonance Imaging: Assessing sex and racial bias in deep learning-based segmentation 92%
Similar papers in this journal
- Predictive modeling of secondary pulmonary hypertension in left ventricular diastolic dysfunction 94%
- Feasibility of estimation of aortic wave intensity using non-invasive pressure recordings in the absence of flow velocity in man. 93%
- A compact multifunctional model of the rabbit atrioventricular node with dual pathways 93%
Similar papers in this journal
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.