Inactivation of Microorganisms in the Complex Regions of Transvaginal Ultrasound Probes By a UVC-LED Light Based Disinfection System
Yasir, M.; Willcox, M.
Show abstract
Endocavity ultrasound transducers, particularly transvaginal ultrasound (TVUS) probes, contain intricate structures such as notches, grooves, lens surfaces, and handle edges that are highly susceptible to microbial contamination yet difficult to disinfect using conventional high-level disinfection (HLD) methods. This study evaluated the efficacy of a novel ultraviolet-C light-emitting diode (UV-C LED) HLD system in eliminating microbial contamination from these complex probe surfaces. Two TVUS probes were sampled from predefined high-risk regions before and after disinfection following clinical use. Probe A was sampled at the top and bottom notches and both sides of the handle, while Probe B was assessed at the lens, edges, and bent groove regions. Microbial contamination was quantified using swab sampling, culture on agar plates, and identification via MALDI-TOF. Environmental sampling of examination and disinfection rooms was also performed. To assess this system robustness, probe sites were repeatedly inoculated with Bacillus subtilis spores and evaluated following UV-C treatment. Before UV-C treatment, contamination rates ranged from 25% to 57% across sampled regions, with microbial loads reaching up to 3.9 log CFU. Identified organisms included Staphylococcus epidermidis, Pseudomonas koreensis, Bacillus cereus, and Propionibacterium spp. Probe sheaths were also predominantly contaminated with Staphylococcus epidermidis., with counts reaching up to 4.3 log CFU, Environmental sampling revealed diverse microbiota, with higher contamination levels in examination rooms compared to disinfection areas. Following 90 seconds of UV-C exposure, no microbial growth was detected on any sampled site, indicating 100% decontamination. Additionally, UV-C treatment achieved >6.7 log reduction of B. subtilis spores across all tested regions. These findings demonstrate that UV-C LED technology provides rapid, effective, and consistent high-level disinfection of complex TVUS probe surfaces, supporting its potential as a rapid and reliable disinfection modality in clinical setting.
Matching journals
The top 10 journals account for 50% of the predicted probability mass.