The Effects of External Laser Positioning Systems for MRI Simulation on Image Quality and Quantitative MRI Values
McCullum, L.; Ding, Y.; Fuller, C. D.; Taylor, B. A.
Show abstract
Background and PurposeMagnetic resonance imaging (MRI) for radiation therapy treatment planning is currently being used in many anatomical sites to better visualize soft tissue landmarks, a technique known as an MRI simulation. A core component of modern MRI simulation configurations are the use of external laser positioning systems (ELPS) to help set up the patient. Though necessary for accurate and reproducible patient setup, the ELPS, if left on during imaging, may interfere negatively with image quality due to leaking electronic noise, of which MRI is sensitive to. It is currently unknown whether this leakage of electronic noise may further affect quantitative values derived from clinically employed relaxometric, diffusion, and fat fraction sequences. Therefore, in this study, we aim to characterize the impact of MRI simulation lasers on general image quality and quantitative imaging accuracy. Materials and MethodsFirst, a cine acquisition was used to visualize the real-time changes in image signal-to-noise ratio (SNR) from when the ELPS was deactivated to activated. To validate this effect quantitatively, the SNR was measured using the American College of Radiology (ACR) recommended protocol in a homogeneous phantom with the integrated body, 18-channel UltraFlex small, 18-channel UltraFlex large, 32-channel spine, and 16-channel shoulder coils. Next, a geometric distortion algorithm was tested in two vendor-provided phantoms while using the integrated body coil and the ACR Large Phantom protocol was tested. Finally, a series of quantitative MRI scans were performed using a CaliberMRI Model 137 Mini Hybrid phantom to validate quantitative T1, T2, and ADC while a Calimetrix PDFF-R2* phantom was used for quantitative PDFF and R2*. All scans were performed with both the ELPS both deactivated and activated. ResultsVisible electronic noise artifacts were seen when using the integrated body coil when the ELPS was activated on the cine acquisition which led to a four-fold decrease in SNR using the ACR protocol. This SNR drop was not seen when using the remaining tested coils. The automatic fiducial detection algorithm was affected negatively by ELPS activation leading to misidentification when identified perfectly with the ELPS deactivated. Degradation in image intensity uniformity, percent signal ghosting, and low contrast object detectability was seen during ACR Large Phantom testing using the 20-channel Head/Neck coil. Concordance across quantitative MRI values was similar when the ELPS was both deactivated and activated while a consistent increase in standard deviation inside the ADC vials was seen when the ELPS was activated. DiscussionThe extra noise induced from the activation of the ELPS during imaging should be avoided due to its potential to unnecessarily increase image noise. This is particularly true when conducting mandatory quality assurance testing for image quality and geometric distortion which utilize the integrated body coil which is most susceptible to ELPS-induced noise. Clear clinical guidelines should be implemented to make this issue known to the MRI technologists, physicists, and other relevant staff using an MRI with a supplementary ELPS for patient alignment. Graphical Abstract O_FIG O_LINKSMALLFIG WIDTH=200 HEIGHT=113 SRC="FIGDIR/small/26347809v1_ufig1.gif" ALT="Figure 1"> View larger version (44K): org.highwire.dtl.DTLVardef@dd725borg.highwire.dtl.DTLVardef@7ed081org.highwire.dtl.DTLVardef@1aac775org.highwire.dtl.DTLVardef@10ce397_HPS_FORMAT_FIGEXP M_FIG C_FIG
Matching journals
The top 3 journals account for 50% of the predicted probability mass.
Similar papers in this journal
Similar papers in this journal
- A Phantom System Designed to Assess the Effects of Membrane Lipids on Water Proton Relaxation 95%
- Non-negative least squares computation for in vivo myelin mapping using simulated multi-echo spin-echo T2 decay data 93%
- MRS-Sim: Open-Source Framework for Simulating Realistic In Vivo-like MR Spectroscopy Data 93%
Similar papers in this journal
- Empirical field mapping for gradient nonlinearity correction of multi-site diffusion weighted MRI 96%
- Increased Brain Volumetric Measurement Precision from Multi-Site 3D T1-weighted 3T Magnetic Resonance Imaging by Correcting Geometric Distortions 95%
- Frequency and Phase Correction of GABA-Edited Magnetic Resonance Spectroscopy using Complex-Valued Convolutional Neural Networks 95%
Similar papers in this journal
- A Detector Block-Pairwise Dead Time Correction Method for Improved Quantitation with a Dedicated BrainPET Scanner 94%
- Patient-derived PixelPrint phantoms for evaluating clinical imaging performance of a deep learning CT reconstruction algorithm 94%
- Reproducible spectral CT thermometry with liver-mimicking phantoms for image-guided thermal ablation 93%
Similar papers in this journal
- Correction of Artifacts Induced by B0 Inhomogeneities in Breast MRI using Reduced Field-of-View Echo-Planar Imaging and Enhanced Reverse Polarity Gradient Method 95%
- Test-retest reproducibility of in vivo magnetization transfer ratio and saturation index in mice at 9.4 Tesla 94%
- The effect of device configuration and patient body composition on image artifact and RF heating of deep brain stimulation devices during MRI at 1.5T and 3T 93%
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.