Echo-planar imaging-based time-of-flight imaging using a modified interleaved flyback trajectory

Bloemer, S.; Stoecker, T.; Stirnberg, R.

2025-10-31 radiology and imaging
10.1101/2025.10.29.25339048
Show abstract

PurposeA modified interleaved flyback (miFB) approach is introduced, designed to mitigate flow artifacts caused by alternating readout polarities in Echo Planar Imaging (EPI), while preserving acquisition efficiency. MethodsWe propose reconstructing odd and even echoes of 3D-EPI separately. To this end, the respective missing lines are acquired in interleaved shots with inverted polarity and an additional precursor gradient. Thereby, high scan efficiency is maintained compared to unsampled flyback gradients. Our miFB approach is additionally combined with gradient moment smoothing and compared to the interleaved dual-echo with acceleration (IDEA) method in phantom and in vivo scans at 7 Tesla. ResultsOur results demonstrate a significant reduction in ghosting and signal dropout using the miFB approach, yielding comparable image quality to non-EPI acquisitions while reducing acquisition time by approximately half. ConclusionThe miFB approach offers a substantial reduction in flow artifacts, allowing for decreased acquisition times in TOF-MRA.