Pulse oximeter performance and skin pigment: comparison of 34 oximeters using current and emerging regulatory frameworks

Hughes, C.; Chen, D.; Law, T.; Bickler, P.; Feiner, J.; Shmuylovich, L.; Behnke, E.; Ortiz, L.; Leeb, G.; Auchus, I.; Negussie, F.; Bisegerwa, R.; Vargas Zamora, R.; Igaga, E.; Moore, K.; Okunlola, O.; Monk, E.; Fernandez, J. L.; Ehie, O.; Wilks, B.; Oyefeso, K.; Schornack, D.; Sendagire, C.; Lipnick, M.

2025-08-13 anesthesia
10.1101/2025.08.11.25332026 medRxiv
Show abstract

BackgroundThe International Organization for Standardization (ISO) and US Food and Drug Administration (FDA) are updating regulations for pulse oximeters to reduce performance disparities linked to skin pigment. We tested common oximeters with current and anticipated regulatory frameworks. We hypothesized that not all oximeters show more positive bias in darkly vs lightly pigmented participants and that few oximeters would pass the anticipated FDA regulations. MethodsWe used a controlled desaturation protocol to test 34 oximeters across arterial oxygen saturations (SaO2) 70-100% in healthy adults. Based on what FDA and ISO had shared at the time of study design, we studied cohort sizes of [≥] 24 with [≥] 25% of participants being darkly pigmented. We used the subjective Monk Skin Tone (MST) scale and the objective individual typology angle (ITA) derived from a spectrophotometer to characterize skin pigment. The root mean square error (ARMS), bias (mean of SpO2 - SaO2 error), and skin pigment differential bias were calculated. Monte Carlo simulation explored potential impacts of participant selection on device passing. ResultsFor cohorts of 24 participants, 28/34 oximeters passed 2017 ISO Standard (ARMS [≤] 4%), 22/34 passed 2013 FDA guidance (ARMS [≤] 3%), 21/34 oximeters passed both ARMS and differential bias criteria for anticipated ISO standards, and 1/34 passed anticipated FDA criteria. More devices passed with cohorts > 24. Eleven oximeters had more positive bias in participants with dark vs. light (dorsal finger) pigmentation across 70-100% SaO2. Eighteen devices could pass or fail depending on cohorts selected for analysis. ConclusionsPulse oximeters show variable performance across manufacturers and models. Notably, only some devices show more positive bias in people with darker skin. Anticipated updates to ISO and FDA frameworks yield strikingly different assessments and require refinement of cohort sizes and differential bias criteria. Whether new guidelines will translate into improved real-world performance or reduced health disparities is yet to be determined.

Published in Anesthesia & Analgesia · not in our set (fewer than 10 published preprints to learn from) · training set

Matching journals

The top 5 journals account for 50% of the predicted probability mass.

50% of probability mass above

"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.