Assessment of repeatability of a clinical automated pupillometer
Diaz, J. C.; Skerswetat, J.; Browne, A. W.
Show abstract
PurposeTo determine the repeatability of a clinical pupillometer in healthy participants with a 5 and 30-minute test-retest break for both Swinging-Flashlight and Low-High Luminance approach. Methods20 healthy participants (mean age: 27 years {+/-}5; 9 females) placed their heads into the devices headrest and fixated a central spot. A Swinging-Flashlight approach evoked a pupillary light reflex, stimulating alternatingly 8 times each eye by a brief diffuse white light flash followed by a continuous measurement of constriction and dilation of the direct and consensual pupils. For the Low-High Luminance setting, continuous measurements of pupil diameters for a 5 second low luminance display followed by a 5 second high luminance white light. Pupillary light reflex parameters for each test and each eye were calculated by the device. Repeatability was investigated after a 5-minute break time and during a control experiment for 21 participants after a 30-minute break for each approach in counterbalanced order using Bland-Altman analysis. ResultsOverall, many parameters for both Swinging-Flashlight and Low-High Luminance approach showed retest biases for all pupil light reflex parameters after a 5-minute test-retest break. These biases were almost completely reduced after a 30-minute break between test and retest for both approaches. The test-retest variabilities as expressed using the Coefficient of Repeatability was reduced after 30-minutes for the majority of the Low-High Luminance results but not for results of the Swinging-Flashlight method. ConclusionsClinical pupillometry includes the Swinging Flashlight Test (SFL) and Low-High Luminance assays. In SFL, many pupillary dynamic metrics showed test-retest bias at a 5-minute interval, whereas relative afferent pupillary defect (RAPD) measurements remained unbiased at both 5 and 30-minute intervals. Similarly, Low-High Luminance testing showed minimal bias after 30 minutes but exhibited bias at the 5-minute interval. Thus, when evaluating multiple pupillary parameters in scientific or clinical settings, RAPD is less susceptible to bias, while other metrics require longer intervals between testing and retesting.
Matching journals
The top 1 journal accounts for 50% of the predicted probability mass.
Similar papers in this journal
- Visual field evaluation using Zippy Adaptive Threshold Algorithm (ZATA) Standard and ZATA Fast in patients with glaucoma and healthy individuals 97%
- Two-Photon Microperimetry: A media opacity-independent retinal function assay 97%
- Relating Standardized Automated Perimetry Performed with Stimulus Sizes III and V in Eyes With Field Loss due to Glaucoma and NAION 96%
Similar papers in this journal
- The evaluation of a web‐based tool for measuring the uncorrected visual acuity and refractive error in keratoconus eyes: a prospective open‐label method comparison study 96%
- Automated 360-degree goniophotography with the NIDEK Gonioscope GS-1 for glaucoma 96%
- A Normative Database of A-Scan Data Using the Heidelberg Spectralis Spectral Domain Optical Coherence Tomography Machine 95%
Similar papers in this journal
Similar papers in this journal
- Automated Expert-level Scleral Spur Detection and Quantitative Biometric Analysis on the ANTERION Anterior Segment OCT System 95%
- Optic nerve tortuosity and displacements during horizontal eye movements in healthy and highly myopic subjects 95%
- Macula structural and vascular differences in glaucoma eyes with and without high axial myopia 95%
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.