Technical Optimization of SyntheticMR for the Head and Neck on a 3T MR-Simulator and 1.5T MR-Linac: A Prospective R-IDEAL Stage 2a Technology Innovation Report
McCullum, L.; Mulder, S. L.; West, N. A.; Scott, H.; Carrasco Rojas, D.; Belal, Z.; Dede, C.; Floyd, W.; Ali, A. M. S.; Mohamed, A. S. R.; Dresner, A.; Subashi, E.; Ma, D.; Stafford, R. J.; Hwang, K.-P.; Fuller, C. D.
Show abstract
ObjectiveThe purpose of this study was to optimize the technical tradeoffs associated with integrating the quantitative maps available from SyntheticMR into the head and neck adaptive radiation oncology workflow. Recent work has begun to investigate SyntheticMR in the adaptive radiation oncology workflow, however no studies have investigated the variation in acquisition parameters and their relationship to the resulting quantitative maps. Filling this gap will facilitate SyntheticMRs translation to the adaptive radiation therapy setting in the head and neck due to the reduced bias and increase repeatability and reproducibility of its generated quantitative relaxometric biomarkers. ApproachThe 2D multi-echo multi-delay (MDME) sequence offered from SyntheticMR was acquired on a MR-Simulation and MR-Linac device in both a phantom and healthy volunteers. Scans were optimized across acceleration factors, slice gaps, acquired voxel sizes, repetition times, echo times, refocusing flip angles, noise-limited gradients, number of slices, and echo train lengths. Quantitative relaxometric T1, T2, and PD maps were tested for accuracy using both correlation and mean absolute bias analysis. Noise profiles were evaluated using the coefficient of variation (CoV) in uniform regions of interest. Main ResultsThe following main findings were reported: (1) noise-limited gradients did not affect the bias or CoV, (2) increasing acceleration factor did not affect bias, but it significantly increased CoV, (3) increasing the number of slices resulted in different significant changes in the quantitative parameters between the MR-Simulator and MR-Linac, (4) reducing the echo train length led to generally reduced bias and CoV, and (5) the repeatability CoV was within previously reported literature values on similar scanners. SignificanceThe behavior of the 2D-MDME sequence from SyntheticMR was characterized across a wide range of acquisition parameters designed for ideal head and neck imaging on both the MR-Simulation and MR-Linac devices in phantom and healthy volunteers. Significant differences existed across several acquisition parameters which should be accounted for when making adjustments for application-specific field-of-views, scan time, and desired ranges of T1, T2, and PD. Application of these findings to the development of efficient head and neck adaptive radiation therapy MRI protocols can enrich the current quantitative biomarker landscape allowing for more informed treatment adaptations.
Matching journals
The top 2 journals account for 50% of the predicted probability mass.
Similar papers in this journal
- Efficient 3D cone trajectory design for improved combined angiographic and perfusion imaging using arterial spin labeling 95%
- Multi-center benchmarking of cervical spinal cord RF coils for 7 T MRI: A traveling spines study 95%
- SNR-Efficient Whole-Brain Pseudo-Continuous Arterial Spin Labeling Perfusion Imaging at 7 Tesla 95%
Similar papers in this journal
- Reproducible spectral CT thermometry with liver-mimicking phantoms for image-guided thermal ablation 95%
- Patient-derived PixelPrint phantoms for evaluating clinical imaging performance of a deep learning CT reconstruction algorithm 95%
- A Detector Block-Pairwise Dead Time Correction Method for Improved Quantitation with a Dedicated BrainPET Scanner 95%
Similar papers in this journal
- MRS-Sim: Open-Source Framework for Simulating Realistic In Vivo-like MR Spectroscopy Data 94%
- A Phantom System Designed to Assess the Effects of Membrane Lipids on Water Proton Relaxation 94%
- Non-negative least squares computation for in vivo myelin mapping using simulated multi-echo spin-echo T2 decay data 94%
Similar papers in this journal
- Quality Assurance Assessment of Intra-Acquisition Diffusion-Weighted and T2-Weighted Magnetic Resonance Imaging Registration and Contour Propagation for Head and Neck Cancer Radiotherapy 95%
- Necessity and Impact of Specialization of Large Foundation Model for Medical Segmentation Tasks 94%
- PSMA-Hornet: fully-automated, multi-target segmentation of healthy organs in PSMA PET/CT images 94%
Similar papers in this journal
- Improved 7 Tesla Transmit Field Homogeneity with Reduced Electromagnetic Power Deposition Using Coupled Tic Tac Toe Antennas 95%
- An in-vivo study of BOLD laminar responses as a function of echo time and static magnetic field strength 94%
- Non-Invasive Perfusion MR Imaging of the Human Brain via Breath-Holding 93%
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.