Differences of blood pressure values between Korotkoff and oscillometric device
Su, H.; jingsong, X.; Xing, C.
Show abstract
SubjectUsing intro-aortic BP as reference,to compare the accuracy of a Korotkoff sound electronic device (Hanvon FY730) and an oscillometric device (Omron HBP-1320T3) in BP measurement, MethodsBP was measured in catheterization coronary angiography. A crossover design was adopted for the using sequence of the two kinds of devices. The average of two BP readings were recorded as the final SBP, DBP and mean arterial pressure (MAP). Non-invasive BP measurement was initiated when the intro-aortic BP curve was stabilized, and the intro-aortic BP curve was simultaneously recorded for 15-20 seconds. BP differences ({Delta}BP) between non-invasive and its relative intro-aortic BP values were calculated as Oscillometric BP - intro-aortic BP difference (Osc-Ao) or Korotkoff sound method - intro-aortic BP difference (Kor-Ao). ResultsUsing intro-aortic BP as the reference, the Korotkoff sound device provided closer readings compared to the oscillometric device.The Kor-Ao SBP was 19.5 {+/-} 17.6 mmHg, but Osc-Ao SBP was 24.8 {+/-} 20.2 mmHg (P < 0.001). Similarly, the MAP difference was smaller for the Korotkoff device (10.5 {+/-} 10.9 mmHg vs. 13.7 {+/-} 11.3 mmHg, P = 0.007). Although the DBP difference was also smaller for the Korotkoff device (6.1 {+/-} 12.3 mmHg vs. 8.1 {+/-} 11.5 mmHg), but no statistical significance (P = 0.147). ConclusionUsing intro-aortic BP reading as reference, the Korotkoff sound device provides more accurate brachial SBP, DBP and MAP values against the oscillometric device.
Matching journals
The top 3 journals account for 50% of the predicted probability mass.
Similar papers in this journal
- Phenylephrine induces relaxation of longitudinal strips from small arteries of goat legs 96%
- Estimating pulse wave velocity from the radial pressure wave using machine learning algorithms 96%
- Preventive training interferes with mRNA-encoding myosin 7 and collagen I expression during pulmonary arterial hypertension 95%
Similar papers in this journal
- Results of the inoperable and operable with aortic valve endocarditis 94%
- Mid-term subclinical myocardial injury detection in patients recovered from COVID-19 according pulmonary lesion severity 94%
- Nomogram-Based Prognostic Model to predict the High blood pressure in Children and Adolescents —— Finding from 342,736 individual in China 94%
Similar papers in this journal
- Safety of anti-thrombotic therapy in acute aortic dissection; single center, observational study 95%
- Detection, Isolation and Quantification of Myocardial Infarct with Four Different Histological Staining Techniques 93%
- Passive Microwave Radiometry (MWR) for diagnostics of COVID-19 lung complications in Kyrgyzstan 92%
Similar papers in this journal
Similar papers in this journal
- Magnesium Sulfate Attenuates Lethality and Oxidative Damage Induced by Different Models of Hypoxia in Mice 91%
- Effects of statins on lipid profile of kidney transplant recipients: a meta-analysis of randomized controlled trials 90%
- Expression of nitric oxide synthase and nitric oxide levels in peripheral blood cells and oxidized low-density lipoprotein levels in saliva as early markers of severe dengue 89%
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.