Cerebellar Ex Vivo Magnetic Resonance Imaging at its Feasibility Limit: Up to 77-Microns Isotropic Resolution using Low-Bandwidth Balanced Steady State Free Precession (LoBa-bSSFP) Sequences and 3T Standard Equipment
Weigel, M.; Dechent, P.; Galbusera, R.; Bahn, E.; Lu, P.-J.; Kappos, L.; Brueck, W.; Stadelmann, C.; Granziera, C.
Show abstract
BackgroundUltra-high-resolution magnetic resonance imaging of the ex vivo brain is increasingly becoming an indispensable tool for studying the morphology and potential pathology of the brain. Despite the important role of the cerebellum in nervous system functions and motor control, as well as its potential damage in neurological diseases, it remains relatively understudied compared to other brain regions. One major reason is the even finer structures. MethodsA balanced steady state free precession approach with receiver bandwidths as low as 50Hz/pixel and long repetition times of 36ms is suggested and optimized, called "LoBa-bSSFP", which enhances the signal-to-noise ratio and alleviates strain on the gradient system for the ultra-high spatial resolutions. A radiofrequency phase cycle scheme is used to reduce potential artifacts. Only 3T MRI standard equipment is utilized for acquisition and basic image reconstruction of the ex vivo brain immersed in perfluoropolyether. ResultsThe presented LoBa-bSSFP approach provides images with very good soft tissue contrast and a detailed visualization of cerebellar morphology. It enables isotropic resolutions of 98-microns for the entire cerebellum, a further refinement allows even up to 77-microns isotropic on a purely clinical MR system. The acquisitions preserved the integrity of the ex vivo cerebellum, so maintaining its connection to the cerebrum and brainstem. ConclusionsOur findings demonstrate the feasibility of employing 3T based LoBa-bSSFP for true ultra-high-resolution ex vivo imaging of the cerebellum, reaching resolutions up to 77-microns isotropic and the potential to reveal subtle microscopic abnormalities of the cerebellar cortex. LoBa-bSSFP may be superior to conventional FLASH sequences in terms of acquisition efficiency and - in some cases - even contrast.
Matching journals
The top 2 journals account for 50% of the predicted probability mass.
Similar papers in this journal
- Combined Angiographic, Structural and Perfusion Radial Imaging using Arterial Spin Labeling 96%
- Vendor-neutral sequences (VENUS) and fully transparent workflows improve inter-vendor reproducibility of quantitative MRI 96%
- Spherical Echo-Planar Time-resolved Imaging (sEPTI) for rapid 3D quantitative T2* and Susceptibility imaging 95%
Similar papers in this journal
- Multiple sclerosis cortical lesion detection with deep learning at ultra-high-field MRI 95%
- 31P Transversal Relaxation Times and Metabolite Concentrations in the Human Brain at 9.4T 95%
- Non-negative least squares computation for in vivo myelin mapping using simulated multi-echo spin-echo T2 decay data 94%
Similar papers in this journal
- An in-vivo study of BOLD laminar responses as a function of echo time and static magnetic field strength 95%
- Temporal signal-to-noise changes in combined multiband- and slice-accelerated echo-planar imaging with a 20- and 64-channel coil 94%
- Improved 7 Tesla Transmit Field Homogeneity with Reduced Electromagnetic Power Deposition Using Coupled Tic Tac Toe Antennas 94%
Similar papers in this journal
- 3D Echo Planar Time-resolved Imaging (3D-EPTI) for ultrafast multi-parametric quantitative MRI 95%
- The dot-compartment revealed? Diffusion MRI with ultra-strong gradients and spherical tensor encoding in the living human brain 95%
- Decoding the microstructural properties of white matter using realistic models 94%
Similar papers in this journal
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.