Back

Novel 10-channel phased-array coil design for carotid wall MRI at 3T

de Buck, M. H. S.; Jezzard, P.; Frost, R.; Randell, C.; Hurst, K.; Choudhury, R. P.; Robson, M. D.; Biasiolli, L.

2022-12-05 radiology and imaging
10.1101/2022.12.01.22283007 medRxiv
Show abstract

BackgroundAccurate assessment of plaque accumulation near the carotid bifurcation is important for the effective prevention and treatment of stroke. However, vessel and plaque delineation using MRI can be limited by low contrast-to-noise ratio (CNR) and long acquisition times. In this work, a novel 10-channel phased-array receive coil design for bilateral imaging of the carotid bifurcation using 3T MRI is proposed. MethodsThe novel 10-channel receive coil was compared to a commercial 4-channel receive coil configuration using data acquired from phantoms and healthy volunteers (N = 9). The relative performance of the coils was assessed, by comparing signal-to-noise ratio (SNR), g-factor noise amplification, and the CNR between vessel wall and lumen using black-blood sequences. Patient data were acquired from 12 atherosclerotic carotid artery disease patients. ResultsThe 10-channel coil consistently provided substantially increased SNR in phantoms (+88 {+/-} 2%) and improved CNR in healthy carotid arteries (+62 {+/-} 11%), or reduced g-factor noise amplification. Patient data showed excellent delineation of atherosclerotic plaque along the length of the carotid bifurcation using the 10-channel coil. ConclusionsThe proposed 10-channel coil design allows for improved visualization of the carotid arteries and the carotid bifurcation and increased parallel imaging acceleration factors.

Matching journals

The top 2 journals account for 50% of the predicted probability mass.

50% of probability mass above

"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.