Validity of Optical Heart Rate Measurement in Commercially Available Wearable Fitness Tracking Devices
Thomas, J.; Doyle, P.; Doyle, A.
Show abstract
BackgroundWearable fitness tracking devices have risen in popularity for athletes and the general population and are increasingly integrated into smartwatch technology. Many devices incorporate optical heart rate (HR) measurement by photoplethysmography which provides data used to monitor and track exercise training intensities, progress, and other health and fitness related parameters. ObjectiveTo determine the validity of optical heart rate measurement in three fitness tracking devices while resting, walking, and running. MethodsTwenty subjects (10 male, 10 female) completed the research study based on the ANSI/CTI standards for physical activity monitoring of heart rate under 4 different conditions: sedentary (SED), treadmill walking (WLK), running (RUN), and dynamic running/walking (DYN). Subjects wore 3 optical heart rate devices: Polar OH1 (OH1) on the right forearm, Apple Watch 4 (AW4) on the right wrist and Garmin Forerunner 945 (FR945) on the left wrist. A Polar H10 (H10), a chest strap device, was the criterion HR measurement device. SED, WLK, and RUN were all 7-minute protocols with 1 minute of standing, 5 minutes of prescribed activity, and 1 final minute of standing. The DYN protocol was a 12-minute protocol with 1 minute of standing, 10 minutes of variable intensity walking and running, and 1 minute of standing. Raw HR data was extracted from each device and temporally aligned with the criterion H10 HR data for analysis. ResultsThe mean absolute deviation (MAD, measured in beats per minute) for the three experimental devices (OH1, AW4, FR945, respectively) for SED was 1.31, 1.33, and 2.03; for WLK was 2.79, 2.58, and 5.19; for RUN were 4.00, 4.29, and 6.51; and for DYN was 2.60, 2.44, and 2.44. The mean absolute percent error (MAPE) for the three experimental devices (OH1, AW4, FR945, respectively) for SED was 1.78%, 1.89%, and 2.81%; for WLK was 3.15%, 3.18%, and 5.93%; for RUN was 3.43%, 3.51%, and 5.25%; and for DYN was 2.05%, 1.95%, and 5.47%. The intraclass correlation for each device across all conditions was .991 (OH1), .984 (AW4), and .697 (FR945). ConclusionsAt rest, and during both steady-state and variable-speed treadmill walking and running, the Polar OH1, Garmin Forerunner 945, and Apple Watch 4 optical HR monitors demonstrated a level of accuracy well within that required by the ANSI/CTA Standard (2018) for physical activity monitoring devices for heart rate measurement (i.e., <10% Mean Absolute Percent Error). Therefore, consumers can have confidence that these devices provide HR data with accuracy that conforms to the performance criteria recommended for consumer electronics.
Matching journals
The top 1 journal accounts for 50% of the predicted probability mass.
Similar papers in this journal
- Comparison of raw accelerometry data from ActiGraph, Apple Watch, Garmin, and Fitbit using a mechanical shaker table 96%
- Validity and reliability of an app-based medical device to empower individuals in evaluating their physical capacities 96%
- Estimation equation of limb lean soft tissue mass in Asian athletes using bioelectrical impedance analysis 94%
Similar papers in this journal
- Ambulatory physiological measures obtained under naturalistic urban mobility conditions have acceptable reliability 95%
- A sweat lactate sensor for detecting anaerobic threshold in patients with heart failure: a prospective clinical trial (LacS-001) 94%
- Effects of Electrocardiograms QRS Detection Algorithms in Heart Rate Variability Metrics 94%
Similar papers in this journal
- Effective Assessments of a Short-duration Poor Posture on Upper Limb Muscle Fatigue before Physical Exercise 94%
- Comparing algorithms for assessing upper limb use with inertial measurement units 93%
- Comparisons of Heart Rate Variability Responses to Head-up Tilt With and Without Abdominal and Lower-Extremity Compression in Healthy Young Individuals: A Randomized Crossover Study 93%
Similar papers in this journal
- Single-session measures of quadriceps neuromuscular function are reliable in healthy females and unaffected by age 93%
- RACLET: the Ramp Above Critical Level Endurance Test to evaluate critical velocity in running. 92%
- A single decreasing ramp friction sprint for torque-cadence relationship assessment during cycling 92%
Similar papers in this journal
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.