In Vivo and Phantom Repeatability of Diffusion-Weighted MRI Sequences on 1.5T MRI-Linear Accelerator (MR-Linac) and MR Simulator Devices for Head and Neck Cancers: Results from a Prospective R-IDEAL Stage 2a Evaluation of Tumor and Normal Tissue Apparent Diffusion Coefficients as Quantitative Imaging Biomarkers
McDonald, B. A.; Salzillo, T.; Mulder, S.; Ahmed, S.; Dresner, A.; Preston, K.; He, R.; Christodouleas, J.; Mohamed, A. S. R.; Philippens, M.; van Houdt, P.; Thorwarth, D.; Wang, J.; Shukla-Dave, A.; Boss, M.; Fuller, C. D.
Show abstract
IntroductionDiffusion-weighted imaging (DWI) on MRI-linear accelerator (MR-linac) systems can potentially be used for monitoring treatment response and adaptive radiotherapy in head and neck cancers (HNC) but requires extensive validation. We perform technical validation to compare six total DWI sequences on an MR-linac and MR simulator (MR sim) in patients, volunteers, and phantoms. MethodsTen human papillomavirus-positive oropharyngeal cancer patients and ten volunteers underwent DWI on a 1.5T MR-linac with three DWI sequences: echo planar imaging (EPI), split acquisition of fast spin echo signals (SPLICE), and turbo spin echo (TSE). Volunteers were also imaged on a 1.5T MR sim with three sequences: EPI, BLADE, and RESOLVE. Participants underwent two scan sessions per device and two repeats of each sequence per session. Repeatability and reproducibility within-subject coefficient of variation (wCV) of mean ADC were calculated for tumors and lymph nodes (patients) and parotid glands (volunteers). Differences in measured ADC values between sequences were quantified using Bland-Altman analysis. ADC bias, repeatability/reproducibility metrics, and SNR were quantified using a phantom. ResultsIn vivo repeatability/reproducibility wCV of mean ADC for parotids were 5.41%/6.72%, 3.83%/8.80%, 5.66%/10.03%, 3.44%/5.70%, 5.04%/5.66%, 4.23%/7.36% for EPIMR-linac, SPLICE, TSE, EPIMR sim, BLADE, RESOLVE. Repeatability/reproducibility wCV for EPIMR-linac, SPLICE, TSE were 9.64%/10.28%, 7.84%/8.96%, 7.60%/11.68% for tumors and 7.80%/9.95%, 7.23%/8.48%, 10.82%/10.44% for nodes. Bland-Altman analysis revealed significant differences between all sequence pairs except BLADE-EPIMR-linac and RESOLVE-SPLICE. All sequences except TSE had phantom ADC biases within {+/-}0.1x10-3 mm2/s for most vials. MR-linac sequences had inconsistent ADC values between different vials with the same known ADC value, indicating spatial inhomogeneities. SNR of b=0 images was 87.3, 180.5, 161.3, 171.0, 171.9, 130.2 for EPIMR-linac, SPLICE, TSE, EPIMR sim, BLADE, RESOLVE. ConclusionMR-linac DWI sequences demonstrate near-comparable performance to MR sim sequences and warrant further clinical validation for treatment response assessment in HNC.
Matching journals
The top 4 journals account for 50% of the predicted probability mass.
Similar papers in this journal
- Multi-center benchmarking of cervical spinal cord RF coils for 7 T MRI: A traveling spines study 95%
- Prospective Motion Correction and Automatic Segmentation of Penetrating Arteries in Phase Contrast MRI at 7 T 94%
- Image Downsampling Expedited Adaptive Least-squares (IDEAL) fitting improves intravoxel incoherent motion (IVIM) analysis in the human kidney 94%
Similar papers in this journal
- Simulated Diagnostic Performance of Ultra-Low-Field MRI: Harnessing Open-Access Datasets to Evaluate Novel Devices 94%
- Increased Brain Volumetric Measurement Precision from Multi-Site 3D T1-weighted 3T Magnetic Resonance Imaging by Correcting Geometric Distortions 94%
- Empirical field mapping for gradient nonlinearity correction of multi-site diffusion weighted MRI 94%
Similar papers in this journal
Similar papers in this journal
- Test-retest reproducibility of in vivo magnetization transfer ratio and saturation index in mice at 9.4 Tesla 93%
- Correction of Artifacts Induced by B0 Inhomogeneities in Breast MRI using Reduced Field-of-View Echo-Planar Imaging and Enhanced Reverse Polarity Gradient Method 92%
- Improved characterization of diffusion in normal and cancerous prostate tissue through optimization of the restriction spectrum imaging signal model 92%
Similar papers in this journal
- A Phantom System Designed to Assess the Effects of Membrane Lipids on Water Proton Relaxation 93%
- Non-negative least squares computation for in vivo myelin mapping using simulated multi-echo spin-echo T2 decay data 93%
- Gradient scheme optimization for PRESS-localized edited MRS using weighted pathway suppression 92%
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.