Diagnostic accuracy of the Panbio COVID-19 Antigen rapid test device for SARS-CoV-2 detection in Kenya, 2021: A field evaluation

Irungu, J. K.; Munyua, P.; Ochieng, C.; Juma, B.; Amoth, P.; Kuria, F.; Kiiru, J.; Makayotto, L.; Abade, A.; Bulterys, M.; Hunsperger, E.; Emukule, G. O.; Onyango, C.; Samandari, T.; Barr, B. A. T.; Akelo, V.; Weyenga, H.; Munywoki, P. K.; Bigogo, G.; Otieno, N. A.; Kisivuli, J. A.; Ochieng, E.; Nyaga, R.; Hull, N.; Herman-Roloff, A.; Aman, R.

2022-05-25 infectious diseases
10.1101/2022.05.23.22275439 medRxiv
Show abstract

BackgroundAccurate and timely diagnosis is essential in limiting the spread of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) infection. Real-time reverse transcription-polymerase chain reaction (rRT-PCR), the reference standard, requires specialized laboratories, costly reagents, and a long turnaround time. Antigen rapid diagnostic tests (Ag RDTs) provide a feasible alternative to rRT-PCR since they are quick, relatively inexpensive, and do not require a laboratory. The WHO requires that Ag RDTs have a sensitivity [&ge;]80% and specificity [&ge;]97%. MethodsThis evaluation was conducted at 11 health facilities in Kenya between March and July 2021. We enrolled persons of any age with respiratory symptoms and asymptomatic contacts of confirmed COVID-19 cases. We collected demographic and clinical information and two nasopharyngeal specimens from each participant for Ag RDT testing and rRT-PCR. We calculated the diagnostic performance of the Panbio Ag RDT against the US Centers for Disease Control and Preventions (CDC) rRT-PCR test. ResultsWe evaluated the Ag RDT in 2,245 individuals where 551 (24.5%, 95% CI: 22.8-26.3%) tested positive by rRT-PCR. Overall sensitivity of the Ag RDT was 46.6% (95% CI: 42.4-50.9%), specificity 98.5% (95% CI: 97.8-99.0%), PPV 90.8% (95% CI: 86.8-93.9%) and NPV 85.0% (95% CI: 83.4-86.6%). Among symptomatic individuals, sensitivity was 60.6% (95% CI: 54.3-66.7%) and specificity was 98.1% (95% CI: 96.7-99.0%). Among asymptomatic individuals, sensitivity was 34.7% (95% CI 29.3-40.4%) and specificity was 98.7% (95% CI: 97.8-99.3%). In persons with onset of symptoms <5 days (594/876, 67.8%), sensitivity was 67.1% (95% CI: 59.2-74.3%), and 53.3% (95% CI: 40.0-66.3%) among those with onset of symptoms >7 days (157/876, 17.9%). The highest sensitivity was 87.0% (95% CI: 80.9-91.8%) in symptomatic individuals with cycle threshold (Ct) values [&le;]30. ConclusionThe overall sensitivity and NPV of the Panbio Ag RDT were much lower than expected. The specificity of the Ag RDT was high and satisfactory; therefore, a positive result may not require confirmation by rRT-PCR. The kit may be useful as a rapid screening tool for only symptomatic patients in high-risk settings with limited access to RT-PCR. A negative result should be interpreted based on clinical and epidemiological information and may require retesting by rRT-PCR.

Matching journals

The top 7 journals account for 50% of the predicted probability mass.

50% of probability mass above

"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.