Diagnostic accuracy of the Panbio COVID-19 Antigen rapid test device for SARS-CoV-2 detection in Kenya, 2021: A field evaluation
Irungu, J. K.; Munyua, P.; Ochieng, C.; Juma, B.; Amoth, P.; Kuria, F.; Kiiru, J.; Makayotto, L.; Abade, A.; Bulterys, M.; Hunsperger, E.; Emukule, G. O.; Onyango, C.; Samandari, T.; Barr, B. A. T.; Akelo, V.; Weyenga, H.; Munywoki, P. K.; Bigogo, G.; Otieno, N. A.; Kisivuli, J. A.; Ochieng, E.; Nyaga, R.; Hull, N.; Herman-Roloff, A.; Aman, R.
Show abstract
BackgroundAccurate and timely diagnosis is essential in limiting the spread of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) infection. Real-time reverse transcription-polymerase chain reaction (rRT-PCR), the reference standard, requires specialized laboratories, costly reagents, and a long turnaround time. Antigen rapid diagnostic tests (Ag RDTs) provide a feasible alternative to rRT-PCR since they are quick, relatively inexpensive, and do not require a laboratory. The WHO requires that Ag RDTs have a sensitivity [≥]80% and specificity [≥]97%. MethodsThis evaluation was conducted at 11 health facilities in Kenya between March and July 2021. We enrolled persons of any age with respiratory symptoms and asymptomatic contacts of confirmed COVID-19 cases. We collected demographic and clinical information and two nasopharyngeal specimens from each participant for Ag RDT testing and rRT-PCR. We calculated the diagnostic performance of the Panbio Ag RDT against the US Centers for Disease Control and Preventions (CDC) rRT-PCR test. ResultsWe evaluated the Ag RDT in 2,245 individuals where 551 (24.5%, 95% CI: 22.8-26.3%) tested positive by rRT-PCR. Overall sensitivity of the Ag RDT was 46.6% (95% CI: 42.4-50.9%), specificity 98.5% (95% CI: 97.8-99.0%), PPV 90.8% (95% CI: 86.8-93.9%) and NPV 85.0% (95% CI: 83.4-86.6%). Among symptomatic individuals, sensitivity was 60.6% (95% CI: 54.3-66.7%) and specificity was 98.1% (95% CI: 96.7-99.0%). Among asymptomatic individuals, sensitivity was 34.7% (95% CI 29.3-40.4%) and specificity was 98.7% (95% CI: 97.8-99.3%). In persons with onset of symptoms <5 days (594/876, 67.8%), sensitivity was 67.1% (95% CI: 59.2-74.3%), and 53.3% (95% CI: 40.0-66.3%) among those with onset of symptoms >7 days (157/876, 17.9%). The highest sensitivity was 87.0% (95% CI: 80.9-91.8%) in symptomatic individuals with cycle threshold (Ct) values [≤]30. ConclusionThe overall sensitivity and NPV of the Panbio Ag RDT were much lower than expected. The specificity of the Ag RDT was high and satisfactory; therefore, a positive result may not require confirmation by rRT-PCR. The kit may be useful as a rapid screening tool for only symptomatic patients in high-risk settings with limited access to RT-PCR. A negative result should be interpreted based on clinical and epidemiological information and may require retesting by rRT-PCR.
Matching journals
The top 7 journals account for 50% of the predicted probability mass.
Similar papers in this journal
- Performance and Implementation Evaluation of the Abbott BinaxNOW Rapid Antigen Test in a High-throughput Drive-through Community Testing Site in Massachusetts 97%
- Comparison of the Accula SARS-CoV-2 Test with a Laboratory-Developed Assay for Detection of SARS-CoV-2 RNA in Clinical Nasopharyngeal Specimens 96%
- Diagnostic accuracy of SARS-CoV-2 rapid antigen detection testing in symptomatic and asymptomatic children in the clinical setting 96%
Similar papers in this journal
- Laboratory and field evaluation of the STANDARD Q and Panbio™ SARS-CoV-2 antigen rapid test in Namibia using nasopharyngeal samples 98%
- Diagnostic accuracy of two commercial SARS-CoV-2 Antigen-detecting rapid tests at the point of care in community-based testing centers 97%
- A prospective diagnostic evaluation of accuracy of self-taken and healthcare worker-taken swabs for rapid COVID-19 testing 96%
Similar papers in this journal
- Evaluation of six commercial SARS-CoV-2 Enzyme-Linked Immunosorbent assays for clinical testing and serosurveillance 94%
- Impact of Nasopharyngeal Specimen Quality on SARS-CoV-2 Test Sensitivity 94%
- Evaluation of a Commercial Culture-free Neutralization Antibody Detection Kit for Severe Acute Respiratory Syndrome-Related Coronavirus-2 and Comparison with an Anti-RBD ELISA Assay 94%
Similar papers in this journal
Similar papers in this journal
- Comparison of Cepheid Xpert Xpress and Abbott ID Now to Roche cobas for the Rapid Detection of SARS-CoV-2 96%
- Clinical evaluation of the molecular-based BD SARS-CoV-2/Flu for the BD MAX™ System 96%
- SARS-CoV-2 patient self-testing with an antigen-detecting rapid test: a head-to-head comparison with professional testing 96%
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.