A comprehensive accuracy assessment of Samsung smartwatch heart rate and heart rate variability

Sarhaddi, F.; Kazemi, K.; Azimi, I.; Cao, R.; Niela-Vil'en, H.; Axelin, A.; Liljeberg, P.; Rahmani, A. M.

2022-05-01 health informatics
10.1101/2022.04.29.22274461 medRxiv
Show abstract

BackgroundPhotoplethysmography (PPG) is a low-cost and easy-to-implement method to measure vital signs, including heart rate (HR) and heart rate variability (HRV). The method is widely used in various wearable devices. For example, Samsung smartwatches are PPG-based open-source wristbands used in remote well-being monitoring and fitness applications. However, PPG is highly susceptible to motion artifacts and environmental noise. A validation study is required to investigate the accuracy of PPG-based wearable devices in free-living conditions. ObjectiveWe evaluate the accuracy of PPG signals - collected by the Samsung Gear Sport smartwatch in free-living conditions - in terms of HR and time-domain and frequency-domain HRV parameters against a medical-grade chest electrocardiogram (ECG) monitor. MethodsWe conducted 24-hours monitoring using a Samsung Gear Sport smartwatch and a Shimmer3 ECG device. The monitoring included 28 participants (14 male and 14 female), where they engaged in their daily routines. We evaluated HR and HRV parameters during the sleep and awake time. The parameters extracted from the smartwatch were compared against the ECG reference. For the comparison, we employed the Pearson correlation coefficient, Bland-Altman plot, and linear regression methods. ResultsWe found a significantly high positive correlation between the smartwatchs and Shimmer ECGs HR, time-domain HRV, LF, and HF and a significant moderate positive correlation between the smartwatchs and shimmer ECGs LF/HF during sleep time. The mean biases of HR, time-domain HRV, and LF/HF were low, while the biases of LF and HF were moderate during sleep. The regression analysis showed low error variances of HR, AVNN, and pNN50, moderate error variances of SDNN, RMSSD, LF, and HF, and high error variances of LF/HF during sleep. During the awake time, there was a significantly high positive correlation of AVNN and a moderate positive correlation of HR, while the other parameters indicated significantly low positive correlations. RMSSD and SDNN showed low mean biases, and the other parameters had moderate mean biases. In addition, AVNN had moderate error variance while the other parameters indicated high error variances. ConclusionThe Samsung smartwatch provides acceptable HR, time-domain HRV, LF, and HF parameters during sleep time. In contrast, during the awake time, AVNN and HR show satisfactory accuracy, and the other HRV parameters have high errors.

Matching journals

The top 6 journals account for 50% of the predicted probability mass.

50% of probability mass above

"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.