Calibration and Evaluation of a Force Measurement Glove for Field-Based Monitoring of Manual Wheelchair Users
Anderson, A. J.; Hooke, A.; Jayaraman, C.; Burns, A.; Fortune, E.; Sosnoff, J. J.; Morrow, M.
Show abstract
Manual wheelchair users exhibit pain related to repetitive and demanding shoulder activities of daily living. Wearable sensing systems like force measurement gloves can provide insights about upper extremity loading in the community and home environments. We calibrated and evaluated the accuracy of a novel force measurement glove with a body-worn data logger. The device was calibrated with loads of 0-800 N applied to the palmar surface of the glove. Calibration conditions were tested that varied the stiffness of the material in the glove, the temperature, and the curvature of the force applicator. Calibration equations from each condition were evaluated by comparing the gloves force prediction with the output of an instrumented wheelchair rim during propulsion and weight relief exercises. The force measurement glove detected 72.7% of 355 propulsion peaks and had a strong linear correlation with the instrumented rim force measurements (r=0.80). The most accurate calibration equation was constructed using data from all conditions, with an RMS force measurement error of 64.7 N and 31.7 N for weight relief exercise and propulsion, respectively. The force measurement glove design described here may serve as a useful tool for detection of loading events and relative magnitude changes.
Matching journals
The top 4 journals account for 50% of the predicted probability mass.
Similar papers in this journal
- Validity and reliability of Kinect v2 for quantifying upper body kinematics during seated reaching 94%
- Development and validation of 2D-LiDAR-based gait analysis instrument and algorithm 94%
- Validity and sensitivity of an inertial measurement unit-driven biomechanical model of motor variability for gait 93%
Similar papers in this journal
- Understanding the Utility of State-Based Haptic Feedback in Tendon-driven Anthropomorphic Prostheses 96%
- Modulation of Arm Swing Frequency and Gait Using Rhythmic Tactile Feedback 95%
- Inertial sensor-based centripetal acceleration as a correlate for lateral margin of stability during walking and turning 95%
Similar papers in this journal
- An inertial sensor-based comprehensive analysis of manual wheelchair user mobility during daily life in people with SCI 96%
- Assessing the validity of the zero-velocity update method for sprinting speeds 95%
- Validation of two-dimensional video-based inference of finger kinematics with pose estimation 95%
Similar papers in this journal
"Similar papers" are the closest papers from that journal in the model's embedding space. They show what the match is built on, but the ranking comes mostly from a classifier over the whole training set, not from these examples alone.