Back

Is the Current N95 Respirator Filtration Efficiency Test Sufficient for Evaluating Protection Against Submicrometer Particles Containing SARS-CoV-2?

2020-05-18 occupational and environmental health Title + abstract only
View on medRxiv
Show abstract

The National Institute of Occupational Safety and Health procedure No. TEB-APR-STP-0059 recommend of measuring the respirator filtration efficiency using sodium chloride aerosol with count median diameter of 75 nm {+/-} 20 nm and geometric standard deviation [≤]1.86. This study showed that this method would overestimate the respirators ability to protect against submicrometer particles. In this study, we converted both mobility diameter and equivalent volume diameter to aerodynamic diameter f...